Technical data

Performance Test Calibration 2
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide 17
[1]
Select NPLC 20 digit resolution.
[2]
The specifications are for 2-wire or 4-wire ohms using the null function in AMM
software. Add 0.2 Ω error if null function is not used in AMM software.
Q = Quick performance verification test points
Short
DC Volts
100 mV ±8 µV
Short 1 V Q ±50 µV
Short 10 V ±0.5 mV
Short 100 V ±5 mV
Short 300 V ±15 mV
Short
2-Wire Ohms
and
4-Wire Ohms
100 Ω ±8 mΩ
[2]
Short 1 kΩ ±50 mΩ
[2]
Short 10 kΩ Q ±500 mΩ
[2]
Short 100 kΩ ±5 Ω
Short 1 MΩ ±50 Ω
Short 10 MΩ ±500 Ω
Short 100 MΩ ±5 kΩ
Table 2 - 2 Zero offset verification test
Input Function
[1]
Range Quick check Error from nominal
1 year