Specifications

11
N9330B Handheld Cable and Antenna Tester Specifications
Test functions
Return loss
SWR
Cable loss
Distance-to-fault (DTF)
Frequency range
25 MHz to 4.0 GHz
Frequency resolution
100 KHz
Output power
0 to –20 dBm
Measurement speed
< 2 second/screen (full span, 521 data point) (CW sweep mode) (3.0 ms/data point, typically)
Number of data points
(Maximum): 521 (Selectable 521, 261, 131)
Return loss
Range 0.00 to 60.00 dB
Accuracy A = 20×log10 (1.1 +10
–(D-RL)/20)
+0.016
(–RL/20)
×10 +10
(–3+RL/20)
)
D Directivity of calibrator
RL Return loss value of DUT
Resolution 0.01 dB
SWR
Range 1.00 to 65.00
Accuracy Same as RL
Resolution 0.01
Cable loss
Range 0.00 to 30.00 dB
Resolution 0.01 dB
Measurement accuracy
> 42 dB corrected directivity after mechanical calibration
> 38 dB corrected directivity after electronic calibration
Distance-to-fault
Vertical range Return loss: 0.00 to 60.00 dB
SWR 1.00 to 65.00
Range
Number of data points - 1) x resolution
Number of data points = 521, 261, or 131
Resolution
Resolution (meter) = (1.5 x 10
8
)x (Vp)/(f2-f1) Hz
Where Vp is the cable's relative propagation velocity
Where f2 is the stop frequency and f1 is start frequency
Markers
6
User storage: Internal user flash memory
Instrument setup storage Up to 15
Trace data storage Up to 200 traces
Screen images storage 10 screens
Support USB memory stick for instrument setup, trace data and screen image storage
General
Display 6.5” 640 x 480 transflective color LCD with adjustable backlight