Installation guide

8
How Does It Work? (continued)
N9340A Handheld Spectrum
Analyzer
Turns the handheld spectrum analyzer
into power meter. Displays power
measurement with its user interface
(UI), which integrates the display of
power measurements.
Keystrokes surrounded by [ ] represent
front-panel hardkeys of the instruments,
while keystrokes surrounded by { }
represent softkeys.
Instructions
1. Power up the N9340A. Connect the U2000 Series USB power sensor to the USB port
of the N9340A (see Figure 13).
2. Press [Mode], use [↑] or [↓] key to select the {Power Meter} then press [ENTER]
(see Figure 14).
Power Meter Menu
Zeroing: Access to zeroing and calibration menu
Meas Disp: Access to measurement display menu and confi gure the displayed
measurement resolution and format
Meas Setup: Set up the relative measurement or set display offsets.
Preset: Preset the USB power sensor
Figure 13 HHSA connect with USB power sensor.
Figure 14 N9340A HHSA turns into a power-meter measurement