User`s guide

Characteristics and Specifications 4
Specification Assumptions
U1731C/U1732C/U1733C User’s Guide 77
Specification Assumptions
Accuracy is given as ±(% of reading + counts of least
significant digit) at 23 °C ± 5 °C, with relative humidity
less than 80% RH.
The measurement performed at the component test socket
and necessary open and short corrections must be done
prior to verifying the instrument’s accuracy.
The accuracy is verified by design and specified type
tests.
SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC) COMPLIANCE
IEC61010-1:2001/EN61010-1:2001 (Second Edition)
IEC 61326-1:2005/EN 61326-1:2006
Canada: ICES/NMB-001:Issue 4, June 2006
Australia/New Zealand: AS/NZS CISPR11:2004
TEMPERATURE COEFFICIENT
0.1 × (specified accuracy) / °C (from –10 °C to 18 °C, or 28 °C to 55 °C)
INPUT PROTECTION
Reset-able over-current protection.
DIMENSIONS (W × H × D)
87 × 184 × 41 mm
WEIGHT
337 grams (with battery)
WARRANTY
Please refer to http://www.agilent.com/go/warranty_terms
Three years for product
Three months for product’s standard accessories, unless otherwise specified
Please take note that for the product, the warranty does not cover:
Damage from contamination
Normal wear and tear of mechanical components
Manuals and standard disposable batteries
CALIBRATION CYCLE
One year