Technical data
VI U1401A User’s and Service Guide
Environmental Conditions
This instrument is designed for indoor use and in an area with low
condensation. The table below shows the general environmental
requirements for this instrument.
Environmental conditions Requirements
Operating temperature Full accuracy from 0 °C to 40 °C
Operating humidity Full accuracy up to 80% R.H. (relative
humidity) for temperature up to 31 °C,
decreasing linearly to 50% R.H. at 40 °C
Storage temperature –20 °C to 60 °C (with battery removed)
Storage humidity 5% to 80% R.H. non-condensing
Altitude Up to 2000 m
Pollution degree Pollution Degree 2
CAUTION
The Handheld Multi-Function Calibrator/Meter complies with the
following safety and EMC requirements.
• IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
• Canada: CAN/CSA-C22.2 No. 61010-1-04
• USA: ANSI/UL 61010-1:2004
• IEC61326-2-1:2005/EN61326-2-1:2006
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
CAUTION
Degradation of some product specifications can occur in the presence of
ambient electromagnetic (EM) fields and noise that are coupled to the
power line or I/O cables of the instrument. The instrument will self-recover
and operate to all specifications when the source of ambient EM field and
noise are removed or when the instrument is protected from the ambient
EM field or when the instrument cabling is shielded from the ambient EM
noise.