Technical data
30 U1401A User’s and Service Guide
2 Calibrator Output Operations
Constant Current Operation
The U1401A can generate a constant current output in the
range of ±25 mA.
To select the constant voltage output function:
1 Turn the rotary switch to any one of the (constant
current output) positions.
2 Press SHIFT to access the shifted operations of the keypad.
The annunciator will appear on the display.
3 Press MODE to cycle through ±25 mA, ±25 mA, and
±25 mA output modes. Select the ±25 mA output
mode for constant output (or steady output, as opposed to
autoscan or autoramp outputs, which will be discussed
under “Memory Generation” on page 31).
• Unlike the autoscan and autoramp modes, there is no
special annunciator on the display to indicate constant
current (CC) operation.
4 With the instrument in standby mode (you should see the
annunciator on the display; if not, press OUTPUT), you
can adjust the amplitude of the output by pressing
and to select the digit to be adjusted, and then
pressing and to adjust the value of the selected
digit.
5 Press OUTPUT to start the source output. The
annunciator will appear on the display.