User`s guide
2 Making Measurements
Measuring AC and DC Signals
44 U1273A/U1273AX User’s Guide
Figure 2-7 AC+DC voltage display
Figure 2-8 AC+DC current display
NOTE
• For better accuracy when measuring the DC offset of an AC voltage,
measure the AC voltage first. Note the AC voltage range, then manually
select a DC voltage range equal to or higher than the AC range. This
procedure improves the accuracy of the DC measurement by ensuring
that the input protection circuits are not activated.
• Press to cycle through the available dual display combinations. See
Appendix B, “Dual Display Combinations Using the Dual Key,” starting
on page 157 to learn more.
• Press to measure the frequency of the AC+DC voltage source.
See “Measuring Frequency” on page 81 to learn more.
Exit
Dual
H
z
%
m
s
L
o
g