User`s guide

2 Making Measurements
Measuring AC and DC Signals
44 U1273A/U1273AX User’s Guide
Figure 2-7 AC+DC voltage display
Figure 2-8 AC+DC current display
For better accuracy when measuring the DC offset of an AC voltage,
measure the AC voltage first. Note the AC voltage range, then manually
select a DC voltage range equal to or higher than the AC range. This
procedure improves the accuracy of the DC measurement by ensuring
that the input protection circuits are not activated.
Press to cycle through the available dual display combinations. See
Appendix B, “Dual Display Combinations Using the Dual Key,” starting
on page 157 to learn more.
Press to measure the frequency of the AC+DC voltage source.
See “Measuring Frequencyon page 81 to learn more.
Exit
Dual
H
z
%
m
s
L
o
g