User`s guide

5 Characteristics and Specifications
Electrical Specifications
150 U1273A/U1273AX User’s Guide
For current measurements
Peak hold specifications
Table 5-15 Frequency sensitivity specifications for current measurements
Input range
[1]
Minimum sensitivity (RMS sine wave)
2 Hz to 30 kHz
300 μA70 μA
3000 μA120 μA
30 mA 1.2 mA
300 mA 12 mA
3 A 0.12 A
10 A 1.2 A
Notes for frequency sensitivity specifications for current measurements:
1 Maximum input for specified accuracy, refer to AC specifications” on page 141.
Table 5-16 Peak hold specifications for DC voltage and current measurements
Signal width Accuracy for DC voltage and current
Single event >1 ms Specified accuracy + 400
Repetitive >250 μs Specified accuracy + 1000