3 Use the arrow keys to change the battery type.
4 Press to save your changes (or press to
discard your changes).
5 Press and hold until the multimeter restarts to return
to normal operation.
Enabling the filter
There are two Filter ( ) options within the design of the
multimeter:
In AC/AC+DC measurement mode, the Filter works as a
low pass filter and attenuates signals with frequencies of
more than 1 kHz
In DC measurement mode, the Filter will block AC signals
Only one of the two filters can be in the signal path at any
point in time. The possible scenarios are:
Only the AC Low Pass Filter is enabled
Only the DC Filter is enabled
No Filter in the signal path
The icon appears when either of the LPF circuits are
enabled. When either LPF is enabled, the measurement
speed (response time) will be impacted.
H
z
%
m
s
L
o
g
S
h
i
f
t
V
i
e
w
E
s
c
Setup
The DC Filter cannot be used when the dual display mode is enabled
where AC and DC voltages are measured.