Enabling smooth mode
Smooth is used to smoothen the refresh rate of the readings
in order to reduce the impact of unexpected noise and to
help you achieve a stable reading.
The smooth refresh rate can be set from 0001 to 9999. The
smooth time is defined as the set value +1. Smooth will be
restarted when the variation count is exceeded, when the
range is changed, or after a multimeter function or feature is
enabled. The variation count is set to the value used for the
AutoHold feature (see “Changing the variation count” on
page 113).
You can enable Smooth by holding while turning on
the multimeter (“Power- on options” on page 13). This
method, however, is temporary and Smooth will be turned
off when you cycle the multimeter’s power. You can
permanently enable Smooth from the setup menu.
To change the smooth refresh rate:
1 Press for more than 1 second to enter the
multimeter’s setup menu.
2 Browse to Menu 5 > SMOOTH, and press edit the
value.
3 Use the arrow keys to change the smooth refresh rate.
Select E to enable the Smooth feature.
4 Press to save your changes (or press to
discard your changes).
5 Press and hold until the multimeter restarts to return
to normal operation.
Parameter Range Default setting
SMOOTH • 0001 to 9999
• D(isabled) or E(nabled)
0009-D(isabled)