User`s guide
3Multimeter Features
Recording Measurement Data (Data Logging)
98 U1273A/U1273AX User’s Guide
Performing manual logs (HAND)
Ensure that HAND is selected as the data logging option in
the multimeter’s setup.
1 Press for more than 1 second to store the present
input signal value.
and the log entry number are displayed at the right
of the display. The display will return to normal after a
short while (around 1 s).
Figure 3-7 Manual log display
2 Repeat step 1 again to save the next input signal value.
The maximum number of readings that can be stored for the
manual log is 100 entries. When all entries are occupied,
H : FULL will be shown when is pressed.
See the “Reviewing Previously Recorded Data (View)” section
later in this manual to review or erase the recorded entries.
Performing interval logs (AUTO)
Ensure that AUTO is selected as the data logging option in
the multimeter’s setup.
The default recording interval duration is 1 second. To
change the recording interval duration, see “Changing the
sample interval duration” on page 114 for more information.