User`s guide
Multimeter Features 3
Capturing Peak Values (Peak)
U1273A/U1273AX User’s Guide 95
To calculate the crest factor:
Crest factor is a measure of signal distortion and is
calculated as a signal’s peak value over its RMS value. This
is an important measurement when looking at power quality
issues. In the measurement example shown below
(Figure 3- 6), the crest factor is calculated as:
Figure 3-6 Peak mode operation
Crest factor
Peak value
True RMS value
-------------------------------------------
2.2669 V
1.6032 V
----------------------
1.414===
Press for more than 1
second to start the Peak Hold
recording session.
P
e
a
k
M
a
x
M
i
n
Press
to reset the
recording
session.
Exit
Dual
Press to switch between the maximum and minimum peak values.
P
e
a
k
M
a
x
M
i
n
Start to capture