Technical data

2 Service and Maintenance
Troubleshooting
52 U1270 Series Service Guide
Troubleshooting
If the instrument fails to operate, check the batteries and
the test leads. Replace them if necessary. If the instrument
still does not function, check the operating procedures in
this manual. When servicing, use only the specified
replacement parts.
The table below will assist you in identifying some basic
malfunctions.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
Any repair or service which is not covered in this manual should only
be performed by qualified personnel.
Table 2 - 1 Operating checklist
Malfunction Identification
No display when powered
ON using the rotary switch
Verify the batteries health and replace
batteries as necessary.
No beeper tone
Verify that the beeper is enabled in the
multimeter’s setup mode.
Failed on current
measurement
Verify the fuses health and replace the fuses
as necessary.
Failed on remote control
Verify the optical side of the IR-USB cable
connected to the multimeter — the Agilent
logo should be facing up.
Verify the baud rate, data bit, and parity
settings in the multimeter’s setup mode.
(Default values are 9600, 8, and none.)
Verify that the driver for the IR-USB interface
is installed.