Technical data
34 Agilent U1251B/U1252B User’s and Service Guide
2 Making Measurements
In the range of 0–500 Ω, the beeper will sound if the
resistance value falls below 10 Ω. For other ranges, the
beeper will sound if the resistance falls below the typical
values indicated in the table below.
The conductance measurement enables the measurement of
very high resistance of up to 100 GΩ. As the high- resistance
readings are susceptible to noise, you can capture the
average readings by using the Dynamic Recording mode.
See Figure 3- 1 on page 47.
Measuring range Beeper sounds when
500.00 ΩΩ < 10
5.0000 kΩΩ < 100
50.000 kΩΩ < 1k
500.00 kΩΩ < 10k
5.0000 MΩΩ < 100k
50.000 MΩΩ < 1M
500.00 MΩΩ < 10M