Technical data

Performance Tests and Calibration 6
Agilent U1251B/U1252B User’s and Service Guide 133
For an adjustment that is successful, the secondary display
will briefly show PASS. An adjustment failure is indicated by
a long beep and a calibration error number appears in the
secondary display. The primary display remains at the
current Cal Item. Check the input value, range, function, and
entered adjustment value to fix the problem. Repeat the
adjustment procedure.
9 Repeat steps 1 through 8 for each adjustment point.
10 Verify the adjustments using the “Performance Verification
Tests” on page 116. Verify the adjustments using
Table 6- 4: