Technical data

Performance Tests and Calibration 6
Agilent U1251B/U1252B User’s and Service Guide 127
To Unsecure the Instrument Without the Security Code
To unsecure the instrument without the correct security
code, follow the steps below.
1 Record the last 4 digits of the multimeter’s serial number.
2 Turn the rotary switch to .
3 Press and simultaneously to enter the
Calibration Security Code entry mode.
The primary display shows 5555 and the secondary
display shows SECUr.
4 Press for more than 1 second to enter the Set
Default Security Code mode. The secondary display shows
SEr.no and the primary display shows 5555.
5 Use the editing keys and to step each
character in the code.
6 Use and to select each character.
7 Set the code, same as the last 4 digits of the instrument’s
serial number.
8 Press
(Save) to confirm the entry.
9 If the 4 digits entered are correct, the secondary display
will show PASS.
NOTE
If you do not have a record of the security code, you can try 1234 (the fac-
tory default code) using the front panel or the remote interface.