Agilent 34980A Mainframe User’s Guide 69
GPIB Interface
Each device on the GPIB (IEEE- 488) interface must have a unique address.
You can set the instrument’s address to any value between 0 and 30.
The address is set to “9” when the instrument is shipped from the factory.
Your computer’s GPIB interface card has its own address. Be sure to
avoid using the computer’s address for any instrument on the interface
bus.
The GPIB address is stored in non- volatile memory, and does not
change when power has been off, after a Factory Reset (
*RST
command), or after an Instrument Preset (
SYSTem:PRESet command).
Front Panel Operation: Utility > REMOTE I/O > GPIB > GPIB ADDRESS
To set the GPIB address, turn the knob (or use the number keypad)
to select the desired address.
Remote Interface Operation: SYSTem:COMMunicate:GPIB:ADDRess
LAN Interface
By default, DHCP is enabled on the instrument, which may enable network
communication over the LAN interface (10BaseT/100BaseTx) without
modifying default settings. However, you may need to change several LAN
configuration parameters, including:
DHCP
IP Address
Auto- IP
Subnet Mask
Default Gateway
Host Name
DNS Server
Domain Name
Front panel and remote interface instructions for setting these parameters
are provided in the following subsections.