96 Agilent 34980A Mainframe User’s Guide
3 Features and Functions
Temperature Measurement Configuration
This section contains information to help you configure the instrument
for making temperature measurements. The table below shows the
thermocouple, RTD, and thermistor types for which the instrument
supports direct measurements.
*
* Using ITS-90 software conversions
** Using ITPS-68 software conversions
Measurement Units
The instrument can report temperature measurements in °C (Celsius), °F
(Fahrenheit), or K (Kelvins). You can mix temperature units on different
channels within the instrument and on the same module.
The
CONFigure and MEASure? commands automatically select °C.
Setting the Mx+B measurement label to °C, °F, or K has no effect on
the temperature measurement units currently selected.
The instrument selects Celsius when the probe type is changed and
after a Factory Reset (
*RST command). An Instrument Preset
(
SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does
not change the units setting.
Front Panel Operation: DMM or Channel (Configure) > TEMPERATURE > UNITS
Remote Interface Operation: UNIT:TEMP {C|F|K}[,(@<ch_list>)]
Thermocouple Types
* RT D Ty p es The r mist o r Ty p es
B, E, J, K, N, R, S, T R
0
= 49 to 2.1 k
= 0.00385 (DIN/IEC 751)
*
= 0.00391**
2.2 k, 5 k, 10 k
(YSI 44000 Series)