Agilent 8702D Lightwave Component Analyzer Installation Guide
© Copyright Agilent Technologies 2000 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under copyright laws. Agilent Part No. 08702-90055 Printed in USA April 2000 Agilent Technologies Lightwave Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, USA (707) 577-1400 Notice. The information contained in this document is subject to change without notice.
Installation at a Glance Installation at a Glance The procedures in this book provide step-by-step instructions for installing the Agilent 8702D. Be sure to verify the Agilent 8702D’s performance Chapter 2, “Automated Verification”, Chapter 3, “Manual Verification”, and Chapter 4, “Performance Tests” provide verification procedures and performance tests. Verification procedures are intended to provide a high level of confidence that the instrument is operating properly.
General Safety Considerations General Safety Considerations This product has been designed and tested in accordance with IEC Publication 61010-1, Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Certification CAUTION Before switching on this instrument, make sure that the line voltage selector switch is set to the line voltage of the power supply and the correct fuse is installed. Assure the supply voltage is in the specified range. CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and 664 respectively.
Assistance Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Service Office.
Contents 1 Installing the Agilent 8702D Step 1. Inspect the Shipment 1-3 Step 2. Set up Static-Safe Workstation 1-4 Step 3. Option 1D5, Connect the Frequency Reference 1-6 Step 4. Check the Fuse and Voltage Selection 1-7 Step 5. Connect the Line-Power Cable 1-9 Step 6. Connect a Keyboard 1-10 Step 7. Turn on the Agilent 8702D 1-11 Step 8. Connect a Printer or Plotter 1-12 Step 9. If You Connect a Printer 1-13 Step 10. If You Connect a Plotter 1-14 Step 11. Set the Clock 1-15 Step 12.
Contents 9. Calibration Coefficients 4-34 10. System Trace Noise (Only for Analyzers without Option 006) 4-40 11. System Trace Noise (Only for Analyzers with Option 006) 4-42 12. Test Port Input Impedance 4-44 13. Test Port Receiver Magnitude Dynamic Accuracy 4-48 14. Test Port Receiver Phase Dynamic Accuracy 4-55 15. Test Port Receiver Magnitude Compression 4-57 16. Test Port Receiver Phase Compression 4-59 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) 4-62 18.
Contents 9. Receiver Input Crosstalk 6-26 10. Receiver Trace Noise 6-33 11. Receiver Input Impedance 6-35 12. Receiver Magnitude Dynamic Accuracy 6-39 13. Receiver Phase Dynamic Accuracy 6-47 14. Receiver Magnitude Compression 6-48 15. Receiver Phase Compression 6-56 16. Source and Receiver Harmonics (Option 002 Only) 6-62 17.
1 Step 1. Inspect the Shipment 1-3 Step 2. Set up Static-Safe Workstation 1-4 Step 3. Option 1D5, Connect the Frequency Reference Step 4. Check the Fuse and Voltage Selection 1-7 Step 5. Connect the Line-Power Cable 1-9 Step 6. Connect a Keyboard 1-10 Step 7. Turn on the Agilent 8702D 1-11 Step 8. Connect a Printer or Plotter 1-12 Step 9. If You Connect a Printer 1-13 Step 10. If You Connect a Plotter 1-14 Step 11. Set the Clock 1-15 Step 12. Check the Operation 1-16 Step 13.
Installing the Agilent 8702D Installing the Agilent 8702D Installing the Agilent 8702D The instructions in this chapter show you how to install your Agilent 8702D. You should be able to finish these procedures in about ten to twenty minutes. Refer to “Specifications and Regulatory Information”, in the Agilent 8702D Reference manual, for information on operating conditions such as temperature. If you should ever need to clean the cabinet, use a damp cloth only.
Installing the Agilent 8702D Step 1. Inspect the Shipment Step 1. Inspect the Shipment 1 Verify that all components ordered have arrived by comparing the shipping forms to the original purchase order. Inspect all shipping containers. If your shipment is damaged or incomplete, save the packing materials and notify both the shipping carrier and the nearest Agilent Technologies service office.
Installing the Agilent 8702D Step 2. Set up Static-Safe Workstation Step 2. Set up Static-Safe Workstation Electrostatic discharge (ESD) can damage or destroy the input circuits of the Agilent 8702D. ESD can also damage or destroy electronic components that you are measuring. All work should be performed at a static-safe work station.
Installing the Agilent 8702D Step 2. Set up Static-Safe Workstation Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 MΩ of isolation from ground. Refer to Table 1-1 for information on ordering staticsafe accessories.
Installing the Agilent 8702D Step 3. Option 1D5, Connect the Frequency Reference Step 3. Option 1D5, Connect the Frequency Reference If your instrument has the optional high stability frequency reference installed, connect the jumper cable on the Agilent 8702D rear panel as shown in Figure 1-2. Figure 1-2.
Installing the Agilent 8702D Step 4. Check the Fuse and Voltage Selection Step 4. Check the Fuse and Voltage Selection 1 Locate the line-input connector on the instrument’s rear panel. 2 Disconnect the line-power cable if it is connected. 3 Use a small flat-blade screwdriver to open the pull-out fuse drawer. WARNING The power cord is connected to internal capacitors that may remain live for 10 seconds after you disconnect the cord from the power supply. Figure 1-3.
Installing the Agilent 8702D Step 4. Check the Fuse and Voltage Selection WARNING For continued protection against fire hazard, replace line fuse only with same type and ratings, (type T 3A/250V for 100/240V operation). The use of other fuses or materials is prohibited. 5 Set the rear panel line-voltage selector to the position that corresponds to the AC power source. Refer to Figure 1-4. Figure 1-4. Line Voltage Selector Table 1-2. Line Power Requirements Power: 115 VAC: 50 WATTS MAX.
Installing the Agilent 8702D Step 5. Connect the Line-Power Cable Step 5. Connect the Line-Power Cable CAUTION Always use the three-prong AC power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage. CAUTION Do not connect ac power until you have verified the line voltage is correct as described in the following paragraphs. Damage to the equipment could result.
Installing the Agilent 8702D Step 6. Connect a Keyboard Step 6. Connect a Keyboard • If you plan to use a keyboard with your Agilent 8702D, connect it now to the instrument’s rear-panel keyboard connector. Figure 1-6.
Installing the Agilent 8702D Step 7. Turn on the Agilent 8702D Step 7. Turn on the Agilent 8702D 1 Press the front-panel LINE key. Refer to Figure 1-7. The front-panel LINE switch disconnects the mains circuits from the mains supply after the EMC filters and before other parts of the instrument.
Installing the Agilent 8702D Step 8. Connect a Printer or Plotter Step 8. Connect a Printer or Plotter 1 Connect a printer or plotter to the corresponding interface port. Refer to Figure 1-6 on page 1-10 for the appropriate rear panel connection. Printer Interface Recommended Cable Parallel Agilent 92284A Serial Agilent 24542G GPIB Agilent 10833A/B/D 2 If you are using the parallel interface, press LOCAL and then PARALLEL until COPY appears.
Installing the Agilent 8702D Step 9. If You Connect a Printer Step 9. If You Connect a Printer 1 Press LOCAL, SET ADDRESSES, PRINTER PORT. 2 Press the key that corresponds to the printer interface: HPIB, PARALLEL (parallel port), or SERIAL (serial port). 3 If you selected HPIB, the GPIB address is active so you can then set the address of your printer. The default GPIB printer address is 1.
Installing the Agilent 8702D Step 10. If You Connect a Plotter Step 10. If You Connect a Plotter 1 Press LOCAL, SET ADDRESSES, PLOTTER PORT. 2 Press the key that corresponds to the plotter interface: HPIB, PARALLEL (parallel port), or SERIAL (serial port). 3 If you selected HPIB, the GPIB address is active so you can then set the address of your plotter. The default GPIB plotter address is 5.
Installing the Agilent 8702D Step 11. Set the Clock Step 11. Set the Clock 1 Press SYSTEM, SET CLOCK to begin setting and activating the time stamp feature so the Agilent 8702D places the time and date on your hardcopies and disk directories. 2 Press the appropriate softkey to set the time and date. 3 Press ROUND SECONDS when the time is exactly as you have set it. 4 Press TIME STAMP so that ON is displayed on the softkey label. Then, press RETURN. Figure 1-8.
Installing the Agilent 8702D Step 12. Check the Operation Step 12. Check the Operation If your instrument is an Agilent 8702D Option 011, skip this step and continue with “Step 13. Check the Operation (Option 011)” on page 1-21. 1 Turn off the LINE power switch. Then, turn back on the LINE power switch. 2 Locate the serial number and configuration options shown on the display. Compare them to the shipment documents. 3 Press PRESET, and observe items shown in Figure 1-9. Figure 1-9.
Installing the Agilent 8702D Step 12. Check the Operation Figure 1-10. Operation Check Connections 5 Press PRESET, SYSTEM, SERVICE MENU, TESTS, EXTERNAL TESTS, EXECUTE TEST. 6 Follow the prompts shown on the display, and then press CONTINUE. 7 Press the up-arrow key, EXECUTE TEST, and then follow the prompts shown on the display. Then press CONTINUE. 8 Connect the equipment as shown in Figure 1-11 and press PRESET.
Installing the Agilent 8702D Step 12. Check the Operation Figure 1-11. Transmission Mode Connections 9 Press CHAN2 to check the forward transmission mode for channel 2. Look at the measurement trace displayed on the Agilent 8702D. It should be similar to the trace shown in Figure 1-12. The Agilent 8702D display shown in Figure 1-12 and Figure 1-13 are Agilent 8702D Option 006 displays (30 kHz to 6 GHz span). Figure 1-12.
Installing the Agilent 8702D Step 12. Check the Operation Figure 1-13. Figure 1-13. Reverse Transmission Display 11 Connect the equipment as shown in Figure 1-14 and press PRESET. Figure 1-14. Transmission Load Match Connections 12 Look at the measurement trace displayed on the Agilent 8702D. It should be similar to the trace displayed in Figure 1-15.
Installing the Agilent 8702D Step 12. Check the Operation Figure 1-15. Forward Transmission Mode 13 Press MEAS, Refl: E S22 REV to check the reverse reflection mode for channel 1. The measurement trace should be similar to Figure 1-16. Figure 1-16. Reverse Reflection Mode 14 Continual with “Step 14. Copy the EEPROM Disk” on page 1-26.
Installing the Agilent 8702D Step 13. Check the Operation (Option 011) Step 13. Check the Operation (Option 011) An Agilent 85047A S-parameter test set must be used when making measurements. 1 Turn off the LINE power switch. Then, turn back on the LINE power switch. 2 Locate the serial number and configuration options shown on the display. Compare them to the shipment documents. 3 Press PRESET, and observe that the analyzer is operating properly as shown in Figure 1-17. Figure 1-17.
Installing the Agilent 8702D Step 13. Check the Operation (Option 011) Figure 1-18. Operation Check Connections 5 Press PRESET, SYSTEM, SERVICE MENU, TESTS, EXTERNAL TESTS, EXECUTE TEST. 6 Follow the prompts shown on the display, and then press CONTINUE. 7 Press the up-arrow key, EXECUTE TEST, and then follow the prompts shown on the display. Then press CONTINUE. 8 Connect the equipment as shown in Figure 1-19 and press PRESET.
Installing the Agilent 8702D Step 13. Check the Operation (Option 011) Figure 1-19. Transmission Mode Connections 9 Press CH2 to check the forward transmission mode for channel 2. Look at the measurement trace displayed on the analyzer. It should be similar to the trace shown in Figure 1-20. The analyzer display shown in Figure 1-20 and Figure 1-21 are Agilent 8702D Option 011, Option 006 displays (30 kHz to 6 GHz span). Figure 1-20.
Installing the Agilent 8702D Step 13. Check the Operation (Option 011) 10 Press MEAS and then Trans: E/E S12 REV to check the reverse transmission mode for channel 2. The measurement trace should be similar to the trace shown in Figure 1-21. Figure 1-21. Reverse Transmission Display 11 Connect the equipment as shown in Figure 1-22 and press PRESET. Figure 1-22. Transmission Load Match Connections 12 Look at the measurement trace displayed on the analyzer.
Installing the Agilent 8702D Step 13. Check the Operation (Option 011) Figure 1-23. Forward Transmission Mode 13 Press MEAS, Refl: E S22 REV to check the reverse reflection mode for channel 1. The measurement trace should be similar to Figure 1-24. Figure 1-24.
Installing the Agilent 8702D Step 14. Copy the EEPROM Disk Step 14. Copy the EEPROM Disk The Agilent 8702D is shipped from the factory with an EEPROM backup disk which is unique to each instrument. The backup disk contains a copy of correction constants which were stored in the instrument’s internal memory at the factory. If your Agilent 8702D should ever be damaged, these correction constants may need to be reinstalled from the disk. Use the following procedure to create a backup copy of the disk.
2 Step 1. Initialization 2-3 Step 2. Measurement Calibration Step 3.
Automated Verification Automated Performance Verification Automated Performance Verification The automated performance verification procedure in this chapter provide a high level of confidence that the Agilent 8702D is working properly. Although this procedure is automated, it does not require a computer. However, you’ll need to locate the Verification Data Disk. If you cannot locate the disk, perform the manual version located in Chapter 3, “Manual Verification”.
Automated Verification Step 1. Initialization How to select the system verification procedure Check to see how the verification kit floppy disk is labeled: • If the disk is labeled Verification Data Disk, proceed with the Agilent 8702D Automated Mode System Verification in this chapter. • If the disk is labeled Verification Data Disc, proceed with the Agilent 8702D Manual Mode System Verification procedure, located in Chapter 3, “Manual Verification”.
Automated Verification Step 2. Measurement Calibration 3 Insert the verification kit disk into the Agilent 8702D’s disk drive. 4 Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK. 5 If you want a printout of the verification data for all the devices, press SYSTEM, SERVICE MENU, TEST OPTIONS, RECORD ON. If you switch on the record function, you cannot switch it off during the verification procedure. 6 Position the paper in the printer so that printing starts at the top of the page.
Automated Verification Step 2. Measurement Calibration Figure 2-2. Connections for Measurement Calibration Standards 5 Press FORWARD:OPEN. 6 When the Agilent 8702D finishes measuring the standard, connect the “short” end of the open/short combination to reference test port 1. 7 Press FORWARD:SHORT. 8 When the Agilent 8702D finishes measuring the standard, connect the 50 ohm termination (supplied in the calibration kit) to reference test port 1. 9 Press FORWARD:LOAD.
Automated Verification Step 3. Device Verification 17 Connect the test port cables as shown Figure 2-3. Figure 2-3. Transmission Calibration Setup 18 Press TRANSMISSION, FWD TRANS THRU. 19 When the Agilent 8702D finishes the measurement, press FWD MATCH THRU. 20 When the Agilent 8702D finishes the measurement, press REV TRANS THRU. 21 When the Agilent 8702D finishes the measurement, press REV MATCH THRU.
Automated Verification Step 3. Device Verification each S-parameter measurement. • If you switched the record function on, the Agilent 8702D measures all S-parameters (magnitude and phase) without pausing. Also, the Agilent 8702D only displays and prints the pass/fail information for the S-parameter measurements that are valid for system verification. Figure 2-4.
Automated Verification Step 3. Device Verification of the verification mismatch to reference test port 1. Figure 2-6. Mismatch Device Verification Setup 7 Press RETURN, TESTS, 30, x1, EXECUTE TEST, CONTINUE. 8 When the Agilent 8702D finishes all the measurements, connect the mismatch verification device, as shown in Figure 2-7. Notice that Port B is now connected to reference test port 1. Figure 2-7. Mismatch Device Verification Setup 9 Press RETURN, TESTS, 31, x1, EXECUTE TEST, CONTINUE.
Automated Verification In Case of Difficulty 10 You have completed the system verification procedure when the Agilent 8702D displays, Ver Def 4 DONE. In Case of Difficulty 1 Inspect all connections. Do not disconnect the cables from the Agilent 8702D test ports. Doing so will invalidate the calibration that you have done earlier. 2 Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN.
3 Step 1. Initialization 3-2 Step 2. Measurement Calibration Step 3.
Manual Verification Manual Performance Verification Manual Performance Verification The manual performance verification procedure in this chapter provides a high level of confidence that the Agilent 8702D is working properly. If you can locate the Agilent 8702D Verification Data Disk, perform the automated version instead, which is documented in Chapter 2, “Automated Verification”. Allow the Agilent 8702D to warm up for one hour before starting this procedure.
Manual Verification Step 1. Initialization Figure 3-1. System Verification Test Setup 2 While the equipment is warming up, review the connector care information in the Agilent 8702D User’s Guide. 3 Set up the system to print the verification data: a Press LOCAL, SYSTEM CONTROLLER. b Press SET ADDRESSES, PRINTER PORT. c Press the softkey that corresponds to your printer port. d If you have difficulty with the printer: • If the interface on your printer is GPIB, verify that the printer address is set to 1.
Manual Verification Step 2. Measurement Calibration CAUTION Do not press PRESET or recall another instrument state. You must use the instrument state that you loaded during the initialization procedure. Step 2. Measurement Calibration 1 Press MENU, POWER, –2, x1 to set the Agilent 8702D test port power to –2 dBm. 2 Press CAL, CAL KITS & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE MENU, FULL 2-PORT. 3 Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 4 Press REFLECTION.
Manual Verification Step 3. Device Verification 18 Connect the test port cables together as shown in Figure 3-2. Figure 3-2. Transmission Calibration Setup 19 Press TRANSMISSION, FWD TRANS THRU. 20 When the Agilent 8702D finishes measuring the standard, press FWD MATCH THRU. 21 When the Agilent 8702D finishes measuring the standard, press REV TRANS THRU. 22 When the Agilent 8702D finishes measuring the standard, press REV MATCH THRU.
Manual Verification Step 3. Device Verification Figure 3-3. Figure 3-3. Connections for the 20 dB Verification Device 3 Press CONTINUE to run the test. 4 Press MENU, POWER, –2, x1, CAL, CORRECTION ON. 5 Press COPY, PRINT MONOCHROME to obtain the graphical representation of the measurement. 6 Press COPY, LIST, PRINT ALL MONOCHROME if you prefer a tabular representation of the measurement. 7 Press SYSTEM, SERVICE MENU, TEST OPTIONS, CONTINUE TEST. Ignore the error message, CAUTION: CALIBRATION REQUIRED.
Manual Verification Step 3. Device Verification tude and phase for the four S-parameters). 10 When the Agilent 8702D displays, Ver Dev 1 FAIL, proceed to the next step only if all the tested S-parameter measurements show a PASS status after you activate error correction in Step 8. Otherwise, refer to “In Case of Difficulty” on page 3-9. 11 Connect the 50 dB attenuator (supplied in the verification kit) as shown in Figure 3-4. Figure 3-4.
Manual Verification Step 3. Device Verification Figure 3-5. Mismatch Device Verification Setup 17 Press TESTS, 30, x1, EXECUTE TEST, CONTINUE. Ignore the error message. 18 Press MENU, POWER, –2, x1, CAL, CORRECTION ON. 19 Repeat Step 5 through Step 8 until you have made all the S-parameter measurements for the verification device.
Manual Verification In Case of Difficulty Figure 3-6. Mismatch Device Verification Setup 22 Press TESTS, 31, x1, EXECUTE TEST, CONTINUE. Ignore the displayed error message. 23 Press MENU, POWER, –2, x1, CAL, CORRECTION ON. 24 Repeat Step 5 through Step 8 until you have made all the S-parameter measurements for the verification device. 25 The test is complete when the Agilent 8702D displays, Ver Def 4 FAIL, and all S-parameter measurements show a PASS status under the same condition mentioned in Step 10.
Manual Verification In Case of Difficulty a Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN. b Use the front panel knob to highlight the calibration you want to recall and press RECALL STATE. c Press MENU, TRIGGER MENU, CONTINUOUS to trigger the continuous sweeps. d Connect the short to reference test port 1. e Press MEAS, Refl: E S11 FWD, MENU, TRIGGER MENU, CONTINUOUS. f Press SCALE REF, SCALE/DIV, .05, x1. g Check to be sure the trace response is 0.00 ±0.05 dB.
4 1. Test Port Output Frequency Range and Accuracy 4-3 2. External Source Mode Frequency Range 4-5 3. Test Port Output Power Accuracy 4-6 4. Test Port Output Power Range and Linearity 4-8 5. Minimum R Channel Level 4-10 6. Test Port Input Noise Floor Level 4-15 7. Test Port Input Frequency Response 4-19 8. Test Port Crosstalk 4-29 9. Calibration Coefficients 4-34 10. System Trace Noise (Only for Analyzers without Option 006) 4-40 11. System Trace Noise (Only for Analyzers with Option 006) 4-42 12.
Performance Tests Manual Performance Tests Manual Performance Tests This chapter contains procedures which test the electrical performance of the Agilent 8702D. If you are using an Agilent 8702D Option 011, perform the tests in Chapter 6, “Performance Tests – Option 011”. Before starting these procedures, allow the instrument to warm up for one hour. Each procedure has a “Performance Test Record” that is located at the end of this chapter.
Performance Tests 1. Test Port Output Frequency Range and Accuracy WARNING Any servicing, adjustment, maintenance, or repair of this product must be performed only by qualified personnel. Repair information provided in the In Case of Difficulty sections of this chapter are provided for qualified service personnel and are intended to be used with the Agilent 8753D Service Guide.
Performance Tests 1. Test Port Output Frequency Range and Accuracy Figure 4-1. Test Port Output Frequency Range and Accuracy Test Setup 2 Press PRESET, MENU, CW FREQ. 3 Press 30, k/m and write the frequency counter reading on the “Performance Test Record.” 4 Repeat Step 3 for each instrument frequency listed in the “Performance Test Record.” In Case of Difficulty 1 If any measured frequency is close to the specification limits, check the time base accuracy of the counter used.
Performance Tests 2. External Source Mode Frequency Range 2. External Source Mode Frequency Range Perform this test to verify that the analyzer’s reference channel, input R, is capable of phase locking to an external CW signal. Required Equipment External Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 83640A Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 Adapter, APC-3.5 (f) to APC-7 . . . . . . . . . . . . . . . . . .
Performance Tests 3. Test Port Output Power Accuracy 5 Check to see if the analyzer is phase locking to the external CW signal: • If the analyzer displays any phase lock error messages, write “unlock” in the “Performance Test Record” for the set CW signal. • If the analyzer does not display any phase lock error messages, write “lock” in the “Performance Test Record” for the set CW signal. 6 On the external source, press CW, 50, MHz/µ. 7 On the analyzer, press 50, M/µ.
Performance Tests 3. Test Port Output Power Accuracy Procedure 1 Zero and calibrate the power meter. For more information of how to perform this task, refer to the power meter operating manual. 2 Connect the equipment as shown in Figure 4-3. Figure 4-3. Source Output Power Accuracy Test Setup 3 Press PRESET.
Performance Tests 4. Test Port Output Power Range and Linearity NOTE The factory preset test port power is 0 dBm. 4 Press MENU, CW FREQ, 300, k/m. Set the calibration factor on the power meter for this CW frequency. 5 Write the power meter reading on the "Performance Test Record." 6 Repeat Step 4 and Step 5 for each CW frequency listed in the "Performance Test Record." In Case of Difficulty 1 Be sure the source power is switched on. Press MENU, POWER. Check the SOURCE PWR softkey, ON should be highlighted.
Performance Tests 4. Test Port Output Power Range and Linearity service manual for more information on how to do this task. 2 On the analyzer, press PRESET, MENU, CW FREQ, 300, k/m. Set the power meter calibration factor for this CW frequency. 3 Connect the equipment as shown in Figure 4-4. Figure 4-4. Test Port Output Power Range and Accuracy Test Setup 4 On the Agilent 438A, press REL. This sets the current power level for relative power measurement. 5 On the analyzer, press MENU, POWER, PWR RANGE MAN.
Performance Tests 5. Minimum R Channel Level 6 Write the power meter reading in the "Results Measured" column on the "Performance Test Record." 7 Calculate the difference between the analyzer test port power (which appears on the analyzer’s display) and the power meter reading. Write the result in the "Power Level Linearity" column on the "Performance Test Record." 8 Repeat Step 5 through Step 7 for the other power levels listed in the "Performance Test Record." 9 Press CW FREQ, 3, G/n.
Performance Tests 5. Minimum R Channel Level Required Equipment For 50Ω Analyzers Adapter, APC-3.5 (m) to APC-7 . . . . . . . . . . . . . . . . . Agilent P/N 1250-1746 Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 For 75Ω Analyzers (Option 075) Minimum Loss Pad, 50Ω to 75Ω . . . . . . . . . . . . . . . . . . . . . . . Agilent 11852B Cable, 50Ω Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781 Adapter, APC-3.
Performance Tests 5. Minimum R Channel Level 8 If the analyzer displays a phase lock error message, continue increasing the test port power until phase lock is achieved. 9 Write the test port power, that is displayed on the analyzer, on the "Performance Test Record." 10 Repeat Step 5 through Step 9 for the other CW frequencies listed in the "Performance Test Record." In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only.
Performance Tests 5. Minimum R Channel Level CAUTION Loop the W8 cable around the A5 sampler as shown in Figure 4-6. Do not push the cable down next to the A11 phase lock assembly. Figure 4-6. Flexible RF Cable Location 2 Using an ohmmeter, verify that the RF cable is not open. In addition, examine both the cable connectors - measure the resistance between the cable center pin and the cable connector and make sure it is not close to zero.
Performance Tests 5. Minimum R Channel Level 3 Check the R sampler by substituting it with the B sampler (A6). a Move cable W8 to the B sampler (A6), as shown in Figure 4-7. Figure 4-7. Connections for Substituting the R Sampler (A4) 4 Connect the equipment as shown in Figure 4-8. Figure 4-8.
Performance Tests 6. Test Port Input Noise Floor Level 5 Repeat the test, but select the B sampler (A6) by pressing MEAS, INPUT PORTS, B in Step 2. Use the following specifications: 300 kHz to 3 GHz 3 GHz to 6 GHz < –27 dBm < –22 dBm 6 If the analyzer fails the test, replace the A11 assembly. 7 Verify that the high/low band adjustments are still within specifications. For more information on how to perform this task, refer to the Agilent 8753D Service Guide.
Performance Tests 6. Test Port Input Noise Floor Level 2 Press PRESET, MENU, POWER, –85, x1, STOP, 3, G/n. 3 Press MEAS, INPUT PORTS, A, TESTPORT 2, FORMAT, LIN MAG, SCALE REF, AUTO SCALE. 4 Press MARKER FCTN, MARKER MODE MENU, STATS ON, MENU, TRIGGER MENU, SINGLE. 5 When the analyzer finishes the sweep, notice the mean value (which appears on the analyzer display). 6 Convert the measured linear magnitude mean value to log magnitude, using this equation.
Performance Tests 6. Test Port Input Noise Floor Level Power (dBm) = 20 * [log10(linear magnitude mean value)] 17 Write this calculated value on the "Performance Test Record." Port 2 Noise Floor Level from 50 kHz to 3 GHz (IF BW= 3 kHz) 18 Press AVG, IF BW, 3, k/m to change the IF bandwidth to 3 kHz. 19 Press MENU, TRIGGER MENU, SINGLE. 20 When the analyzer finishes the sweep, notice the mean value. 21 Convert the measured linear magnitude mean value to log magnitude, using this equation.
Performance Tests 6. Test Port Input Noise Floor Level Port 1 Noise Floor Level for 3 GHz to 6 GHz (IF BW= 10 Hz) 34 Press MEAS, INPUT PORTS, A, TESTPORT 2. 35 Press MENU, TRIGGER MENU, SINGLE. 36 When the analyzer finishes the sweep, notice the mean value. 37 Convert the measured linear magnitude mean value to log magnitude, using this equation. Power (dBm) = 20 * [log10(linear magnitude mean value)] 38 Write this calculated value on the "Performance Test Record.
Performance Tests 7. Test Port Input Frequency Response 7. Test Port Input Frequency Response Perform this test to examine the vector sum of all test setup error vectors in both magnitude and phase change as a function of frequency. Required Equipment For 50Ω Analyzers Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A Cable, APC-7 24-inch . . . . . . . .
Performance Tests 7. Test Port Input Frequency Response Figure 4-10. Setup for Power Meter Calibration on Test Port 1 3 Press PRESET, START, 300, k/m. 4 Only for Analyzers with Option 006: Press STOP, 3, G/n. 5 Press LOCAL, SYSTEM CONTROLLER. 6 Press SET ADDRESSES and POWER MTR until the analyzer shows the correct power meter model. 7 Press ADDRESS: P MTR/HPIB. The default power meter GPIB address is 13. Make sure it is the same as your power meter GPIB address.
Performance Tests 7. Test Port Input Frequency Response 8 Press MENU, NUMBER of POINTS, 51, x1. 9 Press POWER, PWR RANGE MAN to turn the auto power range off. NOTE The analyzer displays the PRm annotation, indicating that the analyzer power range is set to MANUAL. 10 Press PORT POWER to uncouple the test port output power. 11 Press CAL, PWRMTR CAL. 12 Press LOSS/SENSR LISTS, CAL FACTOR SENSOR A.
Performance Tests 7. Test Port Input Frequency Response a From the sensor calibration factor entries menu, press ADD. b Press FREQUENCY, 100, k/m. If you make an entry error, press ← and re-enter the correct value again. c Press CAL FACTOR, 96.4, x1. d Press DONE to terminate the first calibration factor data point entry. e To enter the second cal factor data point, press ADD. f Press FREQUENCY, 300, k/m. g Press CAL FACTOR, 98.4, x1.
Performance Tests 7. Test Port Input Frequency Response Test Port 2 Input Frequency Response from 300 kHz to 3 GHz 16 Connect the equipment as shown in Figure 4-11. Figure 4-11. Test Port 2 Input Frequency Response Test Setup 17 Press MEAS, INPUT PORTS, B. 18 Press SCALE REF, SCALE/DIV, 1, x1. 19 Press MARKER, MARKER 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MIN to put marker 1 at the minimum magnitude location of the trace.
Performance Tests 7. Test Port Input Frequency Response Power Meter Calibration on Port 2 from 300 kHz to 3 GHz 22 Connect the equipment as shown Figure 4-12. Figure 4-12. Setup for Power Meter Calibration on Test Port 2 23 Press MEAS, INPUT PORTS, TESTPORT 2. 24 Press CAL, PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP to start the power meter calibration for test port 2.
Performance Tests 7. Test Port Input Frequency Response 25 When the analyzer displays the message, POWER METER CALIBRATION SWEEP DONE, connect the equipment as shown as in Figure 4-13. Figure 4-13. Test Port 1 Input Frequency Response Test Setup Test Port 1 Input Frequency Response from 300 kHz to 3 GHz 26 Press MEAS, INPUT PORTS, A. 27 Press MARKER, MARKER 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MIN. 28 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MAX.
Performance Tests 7. Test Port Input Frequency Response Figure 4-14. Setup for Power Meter Calibration on Test Port 2 33 Press START, 3, G/n, STOP, 6, G/n. 34 Press CAL, PWRMTR CAL. 35 Press LOSS/SENSR LISTS, CAL FACTOR SENSOR B. Repeat Step 10 to build a calibration factor sensor table for the Agilent 8481A power sensor. 36 Press DONE to exit the sensor calibration factor entries menu. 37 To select the Agilent 8481A power sensor, press USE SENSOR B.
Performance Tests 7. Test Port Input Frequency Response Test Port 1 Input Frequency Response from 3 GHz to 6 GHz 39 When the analyzer finishes the calibration sweep, connect the equipment as shown in Figure 4-15. Figure 4-15. Setup for Test Port 1 Input Frequency Response 40 Press MEAS, INPUT PORTS, A. 41 Press MARKER, MARKER 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MIN to put marker 1 at the minimum magnitude location of the trace.
Performance Tests 7. Test Port Input Frequency Response Figure 4-16. Setup for Power Meter Calibration on Test Port 1 45 Press MEAS, INPUT PORTS, TESTPORT 1. 46 Press CAL, PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP to start the power meter calibration for output test port 1. Test Port 2 Input Frequency Response from 3 GHz to 6 GHz 47 When the analyzer displays the message, POWER METER CALIBRATION SWEEP DONE, connect the equipment as shown as in Figure 4-17.
Performance Tests 8. Test Port Crosstalk Figure 4-17. Test Port 2 Input Frequency Response Test Setup 48 Press MEAS, INPUT PORTS, B. 49 Press MARKER, MARKER 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MIN. 50 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MAX. 51 Write the marker 1 or marker 2 reading, whichever has the largest magnitude, in the “Performance Test Record.” In Case of Difficulty 1 Be sure you have used the correct power sensor for the frequency range.
Performance Tests 8. Test Port Crosstalk Required Equipment For 50Ω Analyzers Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 For 75Ω Analyzers Calibration Kit, 75Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85036B Procedure 1 Connect the equipment as shown in Figure 4-18. Figure 4-18.
Performance Tests 8. Test Port Crosstalk Crosstalk to Test Port 2 from 300 kHz to 3 GHz 4 Press START, 300, k/m, STOP, 3, G/n. 5 Press MEAS, Trans: E/E S21 FWD. 6 Press SCALE REF, REFERENCE VALUE, –100, x1. 7 Press MENU, TRIGGER MENU, SINGLE. 8 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 9 Write the marker value (which appears on the analyzer display) in the “Performance Test Record.” Crosstalk to Test Port 1 from 300 kHz to 3 GHz 10 Press MEAS, Trans: E/E S12 REV.
Performance Tests 8. Test Port Crosstalk In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests 8. Test Port Crosstalk Figure 4-19.
Performance Tests 9. Calibration Coefficients 9. Calibration Coefficients Perform this procedure to verify the analyzer uncorrected test port characteristics. Required Equipment For 50Ω Analyzers Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 For 75Ω Analyzers Calibration Kit, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests 9. Calibration Coefficients 2 Press PRESET, START, 300, k/m. 3 Press CAL, CAL KITS & STDS, SELECT CAL KIT, CAL KIT: 7mm, RETURN, RETURN, CALIBRATE MENU, FULL 2-PORT. 4 Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 5 Connect the “open” end of the open/short combination (supplied in the calibration kit) to analyzer test port 1. 6 Press REFLECTION, FORWARD:OPEN. 7 Connect the “short” end of the open/short combination to analyzer test port 1. 8 Press FORWARD:SHORT.
Performance Tests 9. Calibration Coefficients Figure 4-21. Transmission Calibration Test Setup 19 Press TRANSMISSION, FWD TRANS THRU. 20 When the analyzer finishes the measurement, press FWD MATCH THRU. 21 When the analyzer finishes the measurement, press REV TRANS THRU. 22 After the analyzer finishes the measurement, press REV MATCH THRU. 23 After the analyzer finishes the measurement, press STANDARDS DONE, DONE 2PORT CAL.
Performance Tests 9. Calibration Coefficients Transmission Tracking (Forward) Calibration Coefficient 31 Press SYSTEM, SERVICE MENU, TESTS, 37, x1, EXECUTE TEST. 32 When the analyzer finishes the test, repeat Step 25 through Step 28. Reflection Tracking (Forward) Calibration Coefficient 33 Press SYSTEM, SERVICE MENU, TESTS, 34, x1, EXECUTE TEST. 34 When the analyzer finishes the test, repeat Step 25 through Step 28.
Performance Tests 9. Calibration Coefficients 40 Press PRESET, START, 300, k/m. 41 Press CAL, CAL KITS & STDS, SELECT CAL KIT, CAL KIT: 7mm, RETURN, RETURN, CALIBRATE MENU, FULL 2-PORT. 42 Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 43 Connect the “open” end of the open/short combination (supplied in the calibration kit) to reference test port 1. 44 Press REFLECTION, FORWARD:OPEN. 45 Connect the “short” end of the open/short combination to reference test port 1. 46 Press FORWARD:SHORT.
Performance Tests 9. Calibration Coefficients Figure 4-23. Transmission Calibration Test Setup 57 Press TRANSMISSION, FWD TRANS THRU. 58 When the analyzer finishes the measurement, press FWD MATCH THRU. 59 When the analyzer finishes the measurement, press REV TRANS THRU. 60 When the analyzer finishes the measurement, press REV MATCH THRU. 61 When the analyzer displays, PRESS ’DONE’ IF FINISHED WITH STD(s), press STANDARDS DONE, DONE 2-PORT CAL.
Performance Tests 10. System Trace Noise (Only for Analyzers without Option 006) Source Match (Reverse) Calibration Coefficient 69 Press SYSTEM, SERVICE MENU, TESTS, 39, x1, EXECUTE TEST. 70 When the analyzer finishes the test, repeat Step 63 through Step 66. Reflection Tracking (Reverse) Calibration Coefficient 71 Press SYSTEM, SERVICE MENU, TESTS, 40, x1, EXECUTE TEST. 72 When the analyzer finishes the test, repeat Step 63 through Step 66. 10.
Performance Tests 10. System Trace Noise (Only for Analyzers without Option 006) 2 Press PRESET, MENU, POWER, 5, x1. 3 Press RETURN, CW FREQ, 3, G/n, NUMBER of POINTS, 1601, x1. 4 Press MARKER FCTN, MARKER MODE MENU, STATS ON to activate the instrument’s statistic feature. System Trace Noise for A/R Magnitude 5 Press MEAS, Trans: E/E S12 REV. 6 Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 7 When the analyzer displays the “Hld” annotation, press SCALE REF, AUTO SCALE. 8 Write the s.
Performance Tests 11. System Trace Noise (Only for Analyzers with Option 006) In Case of Difficulty 1 Perform the “ADC Offset Correction Constants” procedure, located in the “Adjustments and Correction Constants” chapter of the Agilent 8753D Service Guide. 2 Repeat this performance test. 3 Suspect the A10 Digital IF board assembly if the analyzer still fails the test. 11.
Performance Tests 11. System Trace Noise (Only for Analyzers with Option 006) 3 Press MARKER FCTN, MARKER MODE MENU, STATS ON to activate the instrument’s statistic feature. System Trace Noise for A/R Magnitude from 30 kHz to 3 GHz 4 Press MEAS, Trans: E/E S12 REV. 5 Press MENU, CW FREQ, 3, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 6 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE. 7 Write the s.
Performance Tests 12. Test Port Input Impedance display, on the “Performance Test Record.” System Trace Noise for B/R Magnitude from 3 GHz to 6 GHz 21 Press MENU, CW FREQ, 6, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 22 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE. 23 Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.
Performance Tests 12. Test Port Input Impedance Required Equipment For 50Ω Analyzers Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B For 75Ω Analyzers Cable, 75Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408 Calibration Kit, 75Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests 12. Test Port Input Impedance 8 When the analyzer displays the prompt, CONNECT STD THEN PRESS KEY TO MEASURE, connect a short to reference test port 1. 9 Press FORWARD: SHORT. 10 At the prompt, connect a load to reference test port 1. 11 Press FORWARD: LOAD. 12 When the analyzer displays, ’DONE’ IF FINISHED WITH CAL, press DONE 1-PORT CAL. 13 Press SAVE/RECALL, SAVE STATE. 14 Connect the equipment as shown in Figure 4-27. Figure 4-27.
Performance Tests 12. Test Port Input Impedance Figure 4-28. S22 1-Port Cal Test Setup 18 Press CAL, CALIBRATE MENU, S22 1-PORT. 19 At the prompt, connect an open to reference test port 2, as shown in Figure 4-28. 20 Press REVERSE: OPEN. 21 When the analyzer displays the prompt, CONNECT STD THEN PRESS KEY TO MEASURE, connect a short to reference test port 2. 22 Press REVERSE: SHORT. 23 At the prompt, connect a load to reference test port 2. 24 Press REVERSE: LOAD.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy Figure 4-29. Test Port 1 Input Impedance Test Setup 28 Press MARKER to activate the analyzer’s marker 1. Use the front panel knob to locate the maximum value of the data trace for each of the frequency ranges listed in the “Performance Test Record.” 29 Write the maximum values on the “Performance Test Record.” In Case of Difficulty 1 Suspect the A10 digital IF board assembly if the analyzer fails both test port tests.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy Figure 4-30. Receiver Magnitude Dynamic Accuracy Required Equipment Measuring Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8902A Step Attenuator, 110 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8496A Adapter (2), APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . Agilent 11524A Adapter, Type-N (m) to Type-N (m) . . . . . . . . . . . . .
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy Figure 4-31.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy 5 On the Agilent 8702D: a Press MEAS, INPUT PORTS, B, MENU, POWER, –4, x1. b Using the analyzer front panel knob, adjust the test port power for a reading of –20.000 dBm on the Agilent 8902A. c Wait for the reading to stabilize. NOTE Ignore the UNCAL displayed on the Agilent 8902A. This indicates that you are making an approximate power level measurement.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy NOTE When the Agilent 8902A displays the “RECAL” annunciation, press CALIBRATE. Test Port 1 Magnitude Dynamic Accuracy 15 Set the Agilent 8496A to 10 dB. 16 On the Agilent 8702D, press CONTINUOUS, MEAS, INPUT PORTS, TESTPORT 2, A. 17 Connect the equipment as shown in Figure 4-32.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy Figure 4-32. Test Port 1 Magnitude Dynamic Accuracy Test Setup 18 On the Agilent 8702D: a Press MENU, POWER. b Using the analyzer front panel knob, adjust the test port power for a reading of 0.000 dB on the Agilent 8902A. c Wait for the reading to stabilize.
Performance Tests 13. Test Port Receiver Magnitude Dynamic Accuracy 19 On the Agilent 8902A, press the blue shift key and SET REF. 20 On the Agilent 8702D: a Press CAL, CALIBRATE MENU, RESPONSE, THRU. b After the analyzer finishes the measurement, press DONE: RESPONSE. c Press SAVE/RECALL, SAVE STATE. 21 Set the Agilent 8496A to 0 dB. 22 On the Agilent 8702D, press MENU, TRIGGER MENU, SINGLE. 23 Write the mean value in the “Test Port” column of the “Performance Test Record.
Performance Tests 14. Test Port Receiver Phase Dynamic Accuracy In Case of Difficulty 1 If the analyzer fails the test at all power levels, be sure you followed the recommended attenuator settings as listed in the “Performance Test Record.” Repeat this performance test. 2 If both test port measured values are out of specifications: a Recalibrate the Agilent 8902A. b Press CLEAR on the Agilent 8902A. c Repeat this performance test.
Performance Tests 14. Test Port Receiver Phase Dynamic Accuracy Figure 4-33. Dynamic Accuracy (Phase) Procedure Test Port 2 Phase Dynamic Accuracy 1 Refer to the “Test Port Receiver Magnitude Dynamic Accuracy” portion of the “Performance Test Record,” for the values in the “Calculated Value” column. 2 Copy the values in the “Magnitude Value” column to the “Calculated Value” column of the “Test Port Receiver Phase Dynamic Accuracy” portion of the “Performance Test Record.
Performance Tests 15. Test Port Receiver Magnitude Compression 7 For every test port input power level, calculate the receiver phase dynamic accuracy using the following formula: Phase Dynamic Accuracy = SIN –1 (B) 8 Write the value in the “Results Value” column of the “Performance Test Record.” Test Port 1 Phase Dynamic Accuracy 9 Repeat Step 1 through Step 8. 15.
Performance Tests 15. Test Port Receiver Magnitude Compression 2 Press PRESET, MEAS, Trans: E/E S21 FWD. 3 Press AVG, IF BW, 10, x1. 4 Press MENU, CW FREQ, 50, M/µ. 5 Press SWEEP TYPE MENU, POWER SWEEP, START, –10, x1. 6 Press MENU, TRIGGER MENU, SINGLE. 7 At the end of the sweep, press SCALE REF, AUTO SCALE. 8 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 9 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MIN. 10 Press MARKER, ∆MODE MENU, ∆REF= 1.
Performance Tests 16. Test Port Receiver Phase Compression SEARCH: MIN. 25 Write the absolute value of the marker 2 reading in the “Measured Value” column of the “Performance Test Record.” 26 Repeat Step 19 through Step 25 for the other CW frequencies listed for Port 1 in the “Performance Test Record.” In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
Performance Tests 16. Test Port Receiver Phase Compression For 75Ω Analyzers Cable, 75Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408 Procedure Test Port 2 Phase Compression 1 Connect the equipment as shown in Figure 4-35. Figure 4-35. Test Port Phase Compression Test Setup 2 Press PRESET, MEAS, Trans: E/E S21 FWD, FORMAT, PHASE. 3 Press AVG, IF BW, 10, x1. 4 Press MENU, SWEEP TYPE MENU, POWER SWEEP, START, –10, x1. 5 Press MENU, CW FREQ, 50, M/µ.
Performance Tests 16. Test Port Receiver Phase Compression Test Port 1 Phase Compression 13 Press MEAS, Trans: E/E S12 REV, FORMAT, PHASE. 14 Press MENU, CW FREQ, 50, M/µ. 15 Press MENU, TRIGGER MENU, SINGLE. 16 At the end of the sweep, press SCALE REF, AUTO SCALE. 17 Press MARKER, ∆ MODE MENU, ∆REF= 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 18 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MIN.
Performance Tests 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) Perform this test to determine the spectral purity of the Agilent 8702D input and output test ports. Required Equipment For 50Ω Analyzers Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 Attenuator (2), 20 dB, APC-7 . . . . . . . . . . . . . . . . .
Performance Tests 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) Figure 4-36. Test Port Output Harmonics Test Setup 5 Press MEAS, Trans: E/E S12 REV, INPUT PORTS, A. 6 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the trace. 7 Press SYSTEM, HARMONIC MEAS, HARMONIC SECOND. 8 After one sweep, press SCALE REF, AUTO SCALE to get a better viewing of the trace. 9 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX.
Performance Tests 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) 15 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 16 After one sweep, press SCALE REF, AUTO SCALE. 17 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 18 Write the marker 1 value on the “Performance Test Record.” Port 1 Input Worst Case 2nd Harmonic 19 Connect the equipment as shown in Figure 4-37. Figure 4-37. Receiver Harmonics Test Setup 20 Press PRESET, MENU, POWER, 8, x1.
Performance Tests 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) channel A) 2nd harmonic. Port 1 Input Worst Case 3rd Harmonic 29 Press STOP, 1, G/n to change the stop frequency for measuring the receiver 3rd harmonic. 30 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF. 31 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the trace. 32 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace.
Performance Tests 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) 47 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the trace. 48 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace. 49 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 50 After one sweep, press SCALE REF, AUTO SCALE. 51 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 52 Write the marker 1 value on the “Performance Test Record.” 18.
Performance Tests 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) Figure 4-38. Test Port Output Harmonics Test Setup 5 Press MEAS, Trans: E/E S12 REV, INPUT PORTS, A. 6 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the trace. 7 Press SYSTEM, HARMONIC MEAS, HARMONIC SECOND. 8 After one sweep, press SCALE REF, AUTO SCALE to get a better viewing of the trace. 9 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX.
Performance Tests 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) 14 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace. 15 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 16 After one sweep, press SCALE REF, AUTO SCALE. 17 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 18 Write the marker 1 value on the “Performance Test Record.” Port 1 Input Worst Case 2nd Harmonic 19 Connect the equipment as shown in Figure 4-39. Figure 4-39.
Performance Tests 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) 27 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 28 Write the marker 1 value (which appears on the analyzer display) on the “Performance Test Record.” This is the worst case port 1 input (receiver channel A) 2nd harmonic. Port 1 Input Worst Case 3rd Harmonic 29 Press STOP, 2, G/n to change the stop frequency for measuring the receiver 3rd harmonic. 30 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF.
Performance Tests 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) harmonic. 46 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF. 47 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the trace. 48 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace. 49 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 50 After one sweep, press SCALE REF, AUTO SCALE. 51 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX.
Performance Tests Performance Test Records Performance Test Records The following tables are to record the performance test records for each of the tests. Test 1. Test Port Output Frequency Range and Accuracy Test Frequencies (MHz) Min. (MHz) Results Measured (MHz) Max. (MHz) Measurement Uncertainty (MHz) 0.03 0.029 999 7 ____________________ 0.030 000 3 ± 0.000 000 036 0.3 0.299 997 ____________________ 0.300 003 ± 0.000 000 360 5.0 4.999 950 ____________________ 5.000 050 ± 0.
Performance Tests Performance Test Records Test 2. External Source Mode Frequency Range Test Frequencies (GHz) Results 0.010 ____________________ 0.020 ____________________ 0.100 ____________________ 1.000 ____________________ 2.000 ____________________ 3.000 ____________________ 4.000 ____________________ 5.000 ____________________ 6.
Performance Tests Performance Test Records Test 3. Test Port Output Power Accuracy Test Frequency Test Port Output Power (dBm) 300 kHz 0 ±1 ____________________ ± 0.15 20 MHz 0 ±1 ____________________ ± 0.10 50 MHz 0 ±1 ____________________ ± 0.10 100 MHz 0 ±1 ____________________ ± 0.10 200 MHz 0 ±1 ____________________ ± 0.10 500 MHz 0 ±1 ____________________ ± 0.10 1 GHz 0 ±1 ____________________ ± 0.10 2 GHz 0 ±1 ____________________ ± 0.
Performance Tests Performance Test Records Test 4. Test Port Output Power Range and Linearity (1 of 2) Test Settings Results Measured (dB) Power Level Linearity (dB) Specification (dB) Measurement Uncertainty (dB) CW Frequency = 300 kHz – 15 __________________ __________________ ± 0.2 ± 0.15 – 13 __________________ __________________ ± 0.2 ± 0.15 – 11 __________________ __________________ ± 0.2 ± 0.15 –9 __________________ __________________ ± 0.2 ± 0.
Performance Tests Performance Test Records Test 4. Test Port Output Power Range and Linearity (2 of 2) Test Settings Results Measured (dB) Power Level Linearity (dB) Specification (dB) Measurement Uncertainty (dB) –1 __________________ __________________ ± 0.2 ± 0.15 +1 __________________ __________________ ± 0.2 ± 0.15 +3 __________________ __________________ ± 0.2 ± 0.15 +5 __________________ __________________ ± 0.5 ± 0.15 +7 __________________ __________________ ± 0.5 ± 0.
Performance Tests Performance Test Records Test 5. Minimum R Channel Level CW Frequency Specification (dB) Test Port Power Measurement Uncertainty (dB) 300 kHz < –35 ____________________ ± 1.0 3.29 MHz < –35 ____________________ ± 1.0 3.31 MHz < –35 ____________________ ± 1.0 15.90 MHz < –35 ____________________ ± 1.0 16.10 MHz < –35 ____________________ ± 1.0 30.90 MHz < –35 ____________________ ± 1.0 31.10 MHz < –35 ____________________ ± 1.0 1.
Performance Tests Performance Test Records Test 6.
Performance Tests Performance Test Records Test 8.
Performance Tests Performance Test Records Test 9. Calibration Coefficients (1 of 2) Frequency Range Specification (dB) Directivity 300 kHz – 1.3 GHz ≥ 35 ____________________ ± 0.9 Directivity 1.3 GHz – 3 GHz ≥ 30 ____________________ ± 0.8 3 GHz – 6 GHz ≥ 25 Test Description Measured Value (dB) Measurement Uncertainty (dB) Forward Direction Directivity (Option 006) ± 0.8 ____________________ Forward Direction Source Match 300 kHz – 1.3 GHz ≥ 16 ____________________ ± 0.
Performance Tests Performance Test Records Test 9. Calibration Coefficients (2 of 2) Test Description Frequency Range Specification (dB) Trans. Tracking 300 kHz – 1.3 GHz ± 1.5 ____________________ ± 0.006 Trans. Tracking 1.3 GHz – 3 GHz ± 1.5 ____________________ ± 0.009 Trans. Tracking Option 006) 3 GHz – 6 GHz ± 2.5 Measured Value (dB) Measurement Uncertainty (dB) ± 0.021 ____________________ Forward Direction Load Match 300 kHz – 1.3 GHz ≥ 18 ____________________ ± 0.
Performance Tests Performance Test Records Test 10. System Trace Noise (Only for Analyzers without Option 006) CW Frequency (GHz) Ratio Measured Value Specification Measurement Uncertainty 3 A/R ____________________ < 0.006 dB rms N/A 3 A/R ____________________ < 0.038° rms N/A 3 B/R ____________________ < 0.006 dB rms N/A 3 B/R ____________________ < 0.038° rms N/A Test 11.
Performance Tests Performance Test Records Test 12. Test Port Input Impedance Test Description Test Port Return Loss (dB) Specification (dB) Measurement Uncertainty (dB) 300 kHz – 1.3 GHz Port 2 ____________________ ≥ 18 ± 1.5 1.3 GHz – 3 GHz Port 2 ____________________ ≥ 16 ± 1.5 3 GHz – 6 GHz (Option 006) Port 2 ≥ 14 ± 1.0 300 kHz – 1.3 GHz Port 1 ____________________ ≥ 18 ± 1.5 1.3 GHz – 3 GHz Port 1 ____________________ ≥ 16 ± 1.
Performance Tests Performance Test Records Test 13. Test Port Receiver Magnitude Dynamic Accuracy Test Port Input Power 8496A Setting (dB) Test Port (dB) 8902A (dB) Calculated Value (dB) Spec. (dB) Mea. Uncer. (dB) Test Port 2 – 10 0 _________ _________ _________ ≤ 0.059 ± 0.01 – 20 10 _________ _________ _________ ≤ 0.050 ± 0.02 – 30 20 _________ _________ _________ ≤ 0.050 ± 0.02 – 40 30 _________ _________ _________ ≤ 0.053 ± 0.
Performance Tests Performance Test Records Test 14. Test Port Receiver Phase Dynamic Accuracy Test Port Input Power Calculated Value (dB) A Results Value (degrees) B Spec. (degrees) Mea. Uncer. Test Port 2 – 10 _________ _________ _________ _________ ≤ 0.42 N/A – 20 _________ _________ _________ _________ ≤ 0.35 N/A – 30 _________ _________ _________ _________ ≤ 0.35 N/A – 40 _________ _________ _________ _________ ≤ 0.
Performance Tests Performance Test Records Test 15. Test Port Receiver Magnitude Compression CW Frequency Test Port Measured Value (dB) Specification (dB) Measurement Uncertainty 50 MHz Port 2 ____________________ ≤ 0.45 N/A 1 GHz Port 2 ____________________ ≤ 0.45 N/A 2 GHz Port 2 ____________________ ≤ 0.45 N/A 3 GHz Port 2 ____________________ ≤ 0.45 N/A 4 GHz (Option 006) Port 2 ≤ 0.80 N/A 5 GHz (Option 006) Port 2 ≤ 0.80 N/A 6 GHz (Option 006) Port 2 ≤ 0.
Performance Tests Performance Test Records Test 16. Test Port Receiver Phase Compression CW Frequency Test Port Measured Value (degrees) Specification (degrees) Measurement Uncertainty 50 MHz Port 2 ____________________ ≤ 6° N/A 1 GHz Port 2 ____________________ ≤ 6° N/A 2 GHz Port 2 ____________________ ≤ 6° N/A 3 GHz Port 2 ____________________ ≤ 6° N/A 4 GHz (Option 006) Port 2 ≤ 7.5° N/A 5 GHz (Option 006) Port 2 ≤ 7.5° N/A 6 GHz (Option 006) Port 2 ≤ 7.
Performance Tests Performance Test Records Test 17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) Test Description Specification (dBc) Measurement Value (dBc) Measurement Uncertainty (dB) Test Port Output Harmonics 2nd ≤ 25 ____________________ ± 1.0 3rd ≤ 25 ____________________ ± 1.0 2nd ≤ 15 ____________________ ± 1.0 3rd ≤ 30 ____________________ ± 1.0 2nd ≤ 15 ____________________ ± 1.0 3rd ≤ 30 ____________________ ± 1.
Performance Tests Performance Test Records Test 18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) Test Description Specification (dBc) Measurement Value (dBc) Measurement Uncertainty (dB) Test Port Output Harmonics 2nd ≤ 25 ____________________ ± 3.0 3rd ≤ 25 ____________________ ± 3.0 2nd ≤ 15 ____________________ ± 3.0 3rd ≤ 30 ____________________ ± 3.0 2nd ≤ 15 ____________________ ± 3.0 3rd ≤ 30 ____________________ ± 3.
5 Agilent 8702D Option 011 and Agilent 85046A/47A System Verification 5-4 Step 1. Initialization 5-4 Step 2. Measurement Calibration 5-6 Step 3. Device Verification 5-8 In Case of Difficulty 5-11 Agilent 8702D Option 011 and Agilent 85044A System Verification 5-13 Step 1. Initialization 5-13 Step 2. Measurement Calibration 5-14 Step 3.
Automated Verification – Option 011 Automated Performance Verification for Option 011 Automated Performance Verification for Option 011 The automated performance verification procedures in this chapter provide a high level of confidence that the Agilent 8702D Option 011 is working properly.
Automated Verification – Option 011 Automated Performance Verification for Option 011 How to select the system verification procedure Check to see how the verification kit floppy disk is labeled: • If the disk is labeled Verification Data Disk, proceed with the Agilent 8702D Option 011 Automated Mode System Verification in this chapter. • If the disk is labeled Verification Data Disc, proceed with the Agilent 8702D Manual Mode System Verification procedure in Chapter 3, “Manual Verification”.
Automated Verification – Option 011 Agilent 8702D Option 011 and Agilent 85046A/47A System Verification Agilent 8702D Option 011 and Agilent 85046A/ 47A System Verification In order to run this procedure, the following equipment is required: S-Parameter test set (for analyzers without Option 006) . . . .Agilent 85046A S-Parameter test set (for analyzers with Option 006) . . . . . . .Agilent 85047A Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automated Verification – Option 011 Step 1. Initialization Figure 5-1. System Verification Test Setup 2 While the equipment is warming up, review the connector care information in the Agilent 8702D User’s Guide. 3 Insert the verification kit disk into the analyzer’s disk drive. 4 Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK. 5 If you want a printout of the verification data for all the devices, press SYSTEM, SERVICE MENU, TEST OPTIONS, RECORD ON.
Automated Verification – Option 011 Step 2. Measurement Calibration Step 2. Measurement Calibration 1 Press CAL, CAL KIT & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE MENU, FULL 2-PORT. 2 Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 3 Press REFLECTION. 4 Connect the “open” end of the open/short combination (supplied in the calibration kit) to reference test port 1, as shown in Figure 5-2. Figure 5-2. Connections for Measurement Calibration Standards 5 Press FORWARD:OPEN.
Automated Verification – Option 011 Step 2. Measurement Calibration 7 Press FORWARD:SHORT. 8 When the analyzer displays the message CONNECT STD THEN PRESS KEY TO MEASURE, connect the 50 ohm termination (supplied in the calibration kit) to reference test port 1, as shown in Figure 5-2. 9 Press FORWARD:LOAD. 10 When the analyzer finishes measuring the standard, connect the “open” end of the open/short combination to reference test port 2, as shown in Figure 5-2. 11 Press REVERSE:OPEN.
Automated Verification – Option 011 Step 3. Device Verification 18 Press TRANSMISSION, FWD TRANS THRU. 19 When the analyzer finishes the measurement, press FWD MATCH THRU. 20 When the analyzer finishes the measurement, press REV TRANS THRU. 21 When the analyzer finishes the measurement, press REV MATCH THRU. 22 When the analyzer displays, PRESS ’DONE’ IF FINISHED WITH STD(s), press STANDARDS DONE, DONE 2-PORT CAL.
Automated Verification – Option 011 Step 3. Device Verification 3 Press CONTINUE to run the test: • If you switched the record function off, you have to press CONTINUE after each S-parameter measurement. • If you switched the record function on, the test displays and prints the pass/ fail information for the S-parameter measurements that are valid for system verification.
Automated Verification – Option 011 Step 3. Device Verification 6 When all measurements are complete, replace the attenuator with the verification mismatch, as shown in Figure 5-6. Figure 5-6. Mismatch Device Verification Setup 7 Press RETURN, TESTS, 30, x1, EXECUTE TEST, CONTINUE.
Automated Verification – Option 011 In Case of Difficulty 8 When the analyzer finishes all the measurements, reconnect the mismatch verification device, as shown in Figure 5-7. Figure 5-7. Mismatch Device Verification Setup 9 Press RETURN, TESTS, 31, x1, EXECUTE TEST, CONTINUE. 10 You have completed the system verification procedure when the analyzer displays, TEST 31 Ver Dev 4 PASS. In Case of Difficulty 1 Inspect all connections. Do not disconnect the cables from the analyzer test ports.
Automated Verification – Option 011 In Case of Difficulty 3 If the analyzer still fails the test, check the measurement calibration as follows: a Press PRESET. b Recall the calibration by pressing SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN. c Use the front panel knob to highlight the calibration you want to recall and press RECALL STATE. d Connect the short to reference test port 1. e Press MEAS, Refl: E S11 FWD. f Press MENU, TRIGGER MENU, CONTINUOUS. g Press SCALE REF, AUTO SCALE, SCALE/DIV, .
Automated Verification – Option 011 Agilent 8702D Option 011 and Agilent 85044A System Verification Agilent 8702D Option 011 and Agilent 85044A System Verification In order to run this procedure, the following equipment is required: Transmission/reflection test set . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85044A Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B Verification Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automated Verification – Option 011 Step 2. Measurement Calibration Figure 5-8. System Verification Test Setup 2 While the equipment is warming up, review the connector care information in the Agilent 8702D User’s Guide. 3 Insert the verification kit disk into the analyzer’s disk drive. 4 Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK. 5 If you want a printout of the verification data for all the devices, press SYSTEM, SERVICE MENU, TEST OPTIONS, RECORD ON.
Automated Verification – Option 011 Step 2. Measurement Calibration MENU, ONE-PATH 2-PORT. 2 Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 3 Press REFLECTION. 4 Connect the “open” end of the open/short combination (supplied in the calibration kit) to reference test port 1, as shown in Figure 5-9. Figure 5-9. Connections for Measurement Calibration Standards 5 Press FORWARD:OPEN. The analyzer displays the message WAIT--MEASURING CAL STD.
Automated Verification – Option 011 Step 3. Device Verification 10 When the analyzer finishes measuring the standard, press STANDARDS DONE. 11 Connect the test port cables as shown Figure 5-10. Figure 5-10. Transmission Calibration Setup 12 Press TRANSMISSION, FWD TRANS THRU. 13 When the analyzer finishes the measurement, press FWD MATCH THRU. 14 When the analyzer displays, PRESS ’DONE’ IF FINISHED WITH STD(s), press STANDARDS DONE, DONE 2-PORT CAL.
Automated Verification – Option 011 Step 3. Device Verification as shown in Figure 5-11. Figure 5-11. Connections for the 20 dB Attenuator 3 Press CONTINUE to run the test: • If you switched the record function off, you have to press CONTINUE after each S-parameter measurement. • If you switched the record function on, the test displays and prints the pass/ fail information for the S-parameter measurements that are valid for system verification.
Automated Verification – Option 011 Step 3. Device Verification 4 When the analyzer finishes all the measurements, connect the 50 dB attenuator (supplied in the verification kit), as shown in Figure 5-12. Figure 5-12. Connections for the 50 dB Attenuator 5 Press RETURN, TESTS, 29, x1, EXECUTE TEST, CONTINUE.
Automated Verification – Option 011 Step 3. Device Verification 6 When all measurements are complete, replace the attenuator with the verification mismatch, as shown in Figure 5-13. Figure 5-13. Mismatch Device Verification Setup 7 Press RETURN, TESTS, 30, x1, EXECUTE TEST, CONTINUE.
Automated Verification – Option 011 In Case of Difficulty 8 When the analyzer finishes all the measurements, reconnect the mismatch verification device, as shown in Figure 5-14. Figure 5-14. Mismatch Device Verification Setup 9 Press RETURN, TESTS, 31, x1, EXECUTE TEST, CONTINUE. 10 You have completed the system verification procedure when the analyzer displays, Ver Def 4 PASS. In Case of Difficulty 1 Inspect all connections. Do not disconnect the cables from the analyzer test ports.
Automated Verification – Option 011 In Case of Difficulty c Press RECALL STATE. d Repeat “Step 3. Device Verification” on page 5-16. 3 If the analyzer still fails the test, check the measurement calibration as follows: a Press PRESET. b Recall the calibration by pressing SAVE/RECALL, SELECT DISK, INTERNAL DISK, RETURN. c Use the front panel knob to highlight the calibration you want to recall and press RECALL STATE. d Connect the short to reference test port 1. e Press SCALE REF, SCALE/DIV, .01, x1.
6 1. Source Frequency Range and Accuracy 6-3 2. Source Power Range, Linearity, and Accuracy 6-5 3. Receiver Minimum R Channel Level 6-9 4. Receiver Minimum R Channel Level for External Source Mode 6-12 5. Receiver Channel Noise Floor Level (Only for Analyzers without Option 006) 6-14 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) 6-17 7. Receiver Magnitude Frequency Response 6-21 8. Receiver Phase Frequency Response 6-24 9. Receiver Input Crosstalk 6-26 10.
Performance Tests – Option 011 Manual Performance Tests - Option 011 Manual Performance Tests - Option 011 This chapter contains procedures which test the electrical performance of the Agilent 8702D Option 011. If you are not using an Agilent 8702D Option 011, refer to Chapter 4, “Performance Tests”. Before starting these procedures, allow the instrument to warm up for one hour. Each procedure has a “Performance Test Record” that is located at the end of this chapter.
Performance Tests – Option 011 1. Source Frequency Range and Accuracy WARNING Any servicing, adjustment, maintenance, or repair of this product must be performed only by qualified personnel. Repair information provided in the In Case of Difficulty sections of this chapter are provided for qualified service personnel and are intended to be used with the Agilent 8753D Service Guide.
Performance Tests – Option 011 1. Source Frequency Range and Accuracy Figure 6-1. Source Frequency Range and Accuracy Test Setup 2 Press PRESET, MENU, CW FREQ. 3 Press 300, k/m and write the frequency counter reading on the "Performance Test Record." 4 Repeat Step 3 for each instrument frequency listed in the “Performance Test Record.” In Case of Difficulty 1 If any measured frequency is close to the specification limits, check the time base accuracy of the counter used.
Performance Tests – Option 011 2. Source Power Range, Linearity, and Accuracy 2. Source Power Range, Linearity, and Accuracy Perform this test to verify the analyzer’s power range and linearity at different CW frequencies throughout the analyzer operating frequency range. Required Equipment Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 2. Source Power Range, Linearity, and Accuracy 2 Zero and calibrate the power meter. Also, set the power meter for dBm. Refer to the power meter operating and service manual for more information on how to do this task. 3 Press PRESET, MENU, POWER, 10, x1. 4 Press MENU, CW FREQ, 300, k/m. Set the power meter calibration factor for this CW frequency (except if you are using an Agilent 436A).
Performance Tests – Option 011 2. Source Power Range, Linearity, and Accuracy 9 Write the power meter reading in the "Second Value" column on the "Performance Test Record." 10 Calculate the path loss through the power splitter with this formula: First Value – Second Value = Path Loss 11 Write the result on the "Performance Test Record." 12 Repeat Step 8 through Step 11 for the other CW frequencies listed on the “Performance Test Record.
Performance Tests – Option 011 2. Source Power Range, Linearity, and Accuracy front panel. 16 Press POWER, –5, x1. Write the power meter reading in the “Measured Value” column on the “Performance Test Record”. 17 On the “Performance Test Record”, copy the “Path Loss” value, previously calculated for this CW frequency, into the “Pass Loss” column.
Performance Tests – Option 011 3. Receiver Minimum R Channel Level 31 Press MENU, CW FREQ, 300, k/m. Set the power calibration factor for this CW frequency. 32 On the “Performance Test Record”, copy the “Path Loss” value, previously calculated for this CW frequency, into the “Pass Loss” column. 33 Calculate the calibrated source power level using this formula: Calibrated Power Level = 10 – Loss 34 Write the result in the “Calibrated Power Level” column.
Performance Tests – Option 011 3. Receiver Minimum R Channel Level Required Equipment Adapter, APC-3.5 (m) to APC-7 . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1746 Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779 Attenuator, 20 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 020 Attenuator, 30 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 3. Receiver Minimum R Channel Level c If the analyzer still doesn’t phase lock, continue increasing the source output power until phase lock is achieved. d Write the marker value in the “Performance Test Record”. • Repeat Step 4 through Step 6 for the other CW frequencies listed in the “Performance Test Record”. In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only.
Performance Tests – Option 011 4. Receiver Minimum R Channel Level for External Source Mode 5 Refer to the Agilent 8753D Option 011 Service Guide for more troubleshooting information. 4. Receiver Minimum R Channel Level for External Source Mode Perform this test to verify proper phase lock in the external source mode, as the specified minimum R input level of –25 dBm. Required Equipment External source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 4. Receiver Minimum R Channel Level for External Source Mode 2 On the external source, press PRESET, POWER LEVEL, 4, dB(m), CW, 10, MHz. 3 On the analyzer, press PRESET, MENU, CW FREQ, 10, M/µ, SYSTEM, INSTRUMENT MODE, EXT SOURCE AUTO to set up the analyzer for an external source input to the receiver channel R.
Performance Tests – Option 011 5. Receiver Channel Noise Floor Level (Only for Analyzers without Option 006) In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests – Option 011 5. Receiver Channel Noise Floor Level (Only for Analyzers without Option 006) Figure 6-7. Receiver Channel Noise Floor Level Test Setup 2 Press PRESET, MENU, POWER, –10, x1. Receiver Channel A Noise Floor Level with a 3 kHz IF BW 3 Press MEAS, INPUT PORTS, A, FORMAT, LIN MAG, SCALE REF, AUTO SCALE. 4 Press MARKER FCTN, MARKER MODE MENU, STATS ON, MENU, TRIGGER MENU, SINGLE. 5 When the analyzer finishes the sweep, notice the mean value, which appears on the analyzer display.
Performance Tests – Option 011 5. Receiver Channel Noise Floor Level (Only for Analyzers without Option 006) 11 Use the following equation to convert the linear magnitude mean value to log magnitude: Power (dBm) = 20 * [log10(linear magnitude mean value)] 12 Write this calculated value in the “Performance Test Record” in dBm. Receiver Channel B Noise Floor Level with 10 Hz IF BW 13 Press MEAS, INPUT PORTS, B, FORMAT, LIN MAG. 14 Press MENU, TRIGGER MENU, SINGLE.
Performance Tests – Option 011 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) Perform this test to determine the analyzer receiver channel noise floor levels. Required Equipment Attenuator, 20 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 020 Termination (2), 50Ω . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 908A Cable, 50Ω, Type-N, 24 inch . .
Performance Tests – Option 011 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) 5 When the analyzer finishes the sweep, notice the mean value, which appears on the analyzer display. 6 Use the following equation to convert the linear magnitude mean value to log amplitude: Power (dBm) = 20 * [log10(linear magnitude mean value)] 7 Write this calculated value in the “Performance Test Record” in dBm.
Performance Tests – Option 011 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) Receiver Channel B Noise Floor Level with a 3 kHz IF BW (50 kHz–3 GHz) 18 Press AVG, IF BW, 3, k/m to change the IF bandwidth to 3 kHz. 19 Press MENU, TRIGGER MENU, SINGLE. 20 When the analyzer finishes the sweep, record the mean value, which appears on the analyzer display.
Performance Tests – Option 011 6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006) Receiver Channel A Noise Floor Level with 10 Hz IF BW (3 GHz–6 GHz) 33 Press MEAS, INPUT PORTS, A. 34 Press MENU, TRIGGER MENU, SINGLE. 35 When the analyzer finishes the sweep, record the mean value, which appears on the analyzer display.
Performance Tests – Option 011 7. Receiver Magnitude Frequency Response 7. Receiver Magnitude Frequency Response Perform this test to verify how well the analyzer transfers information from the RF to IF, and how accurately it processes and displays that information. Required Equipment Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 7. Receiver Magnitude Frequency Response 3 Press PRESET, MARKER, MEAS, INPUT PORTS, R, MENU, POWER, 6, x1. 4 Press CW FREQ, 300, k/m. Set the calibration factor on the power meter for this CW frequency. 5 Write the power meter reading on the “Performance Test Record.” 6 Write the marker reading, which appears on the analyzer display, in the “R Input Power” column of the “Performance Test Record.
Performance Tests – Option 011 7. Receiver Magnitude Frequency Response Input B Magnitude Frequency Response 13 Connect the equipment as shown in Figure 6-11. Figure 6-11. Magnitude Frequency Response Test Setup (Receiver Input B) 14 Press MEAS, INPUT PORTS, B, MENU to start measuring the power delivered to receiver input B. 15 Press CW FREQ, 300, k/m. 16 Write the marker reading, which appears on the analyzer display, in the “B Input Power” column of the “Performance Test Record.
Performance Tests – Option 011 8. Receiver Phase Frequency Response NOTE The final results represent the worst case magnitude frequency response for all the analyzer receiver inputs at the selected frequencies. In Case of Difficulty 1 Suspect poor tracking between the ports of the power splitter if all receiver inputs fail. Try to repeat the test with another power splitter. 2 Suspect source drift if receiver inputs A and B fail this test at any frequency.
Performance Tests – Option 011 8. Receiver Phase Frequency Response Figure 6-12. Phase Frequency Response Test Setup Phase Frequency Response from 300 kHz to 3 GHz 2 Press PRESET, MENU, POWER, 10, x1. 3 Only for Analyzers with Option 006: Press STOP, 3, G/n. 4 Press FORMAT, PHASE, MARKER FCTN, MARKER MODE MENU, STATS ON. 5 Press SCALE REF, .6, x1, ELECTRICAL DELAY. Turn the analyzer front panel knob to vary the electrical delay until the trace is in the most linear horizontal position.
Performance Tests – Option 011 9. Receiver Input Crosstalk 12 Press SCALE REF, ELECTRICAL DELAY. Turn the analyzer front panel knob to vary the electrical delay until the trace is in the most linear horizontal position. 13 Press MENU, TRIGGER MENU, SINGLE. Write the p-p value, which appears on the analyzer display, on the “Performance Test Record.” 14 Press CONTINUOUS, MEAS, INPUT PORTS, A/R. 15 Repeat Step 12 and Step 13. 16 Press CONTINUOUS, MEAS, INPUT PORTS, B/R. 17 Repeat Step 12 and Step 13.
Performance Tests – Option 011 9. Receiver Input Crosstalk Required Equipment 2-way Power Splitter (2), 50Ω . . . . . . . . . . . . . . . .Agilent 11667A Option 001 Termination (3), 50Ω . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 908A Attenuator, 20 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 020 RF cable set, 50Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 9. Receiver Input Crosstalk (Only for Option 006 Analyzers) R into A Crosstalk from 3 GHz to 4.5 GHz 9 Press START, 3, G/n, STOP, 4.5, G/n, NUMBER of GROUPS, 5, x1. 10 When the analyzer finishes the number of sweeps, write the marker value, which appears on the analyzer display, on the “Performance Test Record.” (Only for Option 006 Analyzers) R into A Crosstalk from 4.5 GHz to 6 GHz 11 Press START, 4.5, G/n, STOP, 6, G/n, NUMBER of GROUPS, 5, x1.
Performance Tests – Option 011 9. Receiver Input Crosstalk B into A Crosstalk from 300 kHz to 1 GHz 22 Connect the equipment as shown in Figure 6-14. Figure 6-14. B into A Receiver Input Crosstalk Test Setup 23 Press START, 300, k/m, STOP, 1, G/n. 24 Press MEAS, INPUT PORTS, A/B, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 25 When the analyzer finishes the number of sweeps, write the marker value, which appears on the analyzer display, on the “Performance Test Record.
Performance Tests – Option 011 9. Receiver Input Crosstalk (Only for Option 006 Analyzers) B into A Crosstalk from 4.5 GHz to 6 GHz 30 Press START, 4.5, G/n, STOP, 6, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 31 When the analyzer finishes the number of sweeps, write the marker value, which appears on the analyzer display, on the “Performance Test Record.” A into B Crosstalk from 300 kHz to 1 GHz 32 Connect the equipment as shown in Figure 6-15. Figure 6-15.
Performance Tests – Option 011 9. Receiver Input Crosstalk A into B Crosstalk from 1 GHz to 3 GHz 38 Press START, 1, G/n, STOP, 3, G/n. 39 Press MEAS, INPUT PORTS, A/R, AVG, AVERAGING OFF, MENU, TRIGGER MENU, SINGLE. 40 At the end of the sweep, press DISPLAY, DATA →MEMORY, DATA/MEM, MEAS, INPUT PORTS, B/R. 41 Press AVG, AVERAGING FACTOR, 5, x1, AVERAGING ON, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
Performance Tests – Option 011 9. Receiver Input Crosstalk In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests – Option 011 10. Receiver Trace Noise 10. Receiver Trace Noise Perform this test to verify the receiver trace noise on a CW signal in ratio mode. Required Equipment 3-way Power Splitter, 50Ω . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11850C Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 010 RF cable set, 50Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B RF cable, 50Ω, Type-N . . . . . . . .
Performance Tests – Option 011 10. Receiver Trace Noise statistic feature. 5 Write the s.dev (standard deviation) value, which appears on the analyzer display, on the "Performance Test Record." 6 Press MEAS, INPUT PORTS, B/R. Write the s.dev (standard deviation) value on the "Performance Test Record." 7 Press A/B. Write the s.dev (standard deviation) value on the "Performance Test Record." 8 Press FORMAT, PHASE. Write the s.dev (standard deviation) value on the "Performance Test Record.
Performance Tests – Option 011 11. Receiver Input Impedance 11. Receiver Input Impedance Perform this test to verify the return loss of each of the analyzer’s receiver inputs. Required Equipment Test set, 50Ω . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85044A/85046A/85047A 2-way Power Splitter, 50Ω . . . . . . . . . . . . . . . . . . .Agilent 11667A Option 001 Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 010 Termination, 50Ω . . . . . . . . . .
Performance Tests – Option 011 11. Receiver Input Impedance 2 Press PRESET, MENU, NUMBER of POINTS, 1601, x1. 3 Press CAL, CAL KIT & STDS, SELECT CAL KIT, N 50 Ω, RETURN, RETURN, CALIBRATE MENU, S11 1-PORT. 4 Connect an open to the test port cable adapter, by pressing FORWARD: OPENS, OPEN (M), DONE: OPENS. 5 Connect a short to the test port cable adapter, by pressing FORWARD: SHORTS, SHORT (M), DONE: SHORTS.
Performance Tests – Option 011 11. Receiver Input Impedance Figure 6-18. Receiver A Input Impedance Test Setup 14 Press PRESET, MEAS, INPUT PORTS, B/R, MENU, NUMBER of POINTS, 1601, x1. 15 Press CAL, CALIBRATE MENU, S11 1-PORT. Repeat Step 4 through Step 12 for the analyzer receiver input A return loss.
Performance Tests – Option 011 11. Receiver Input Impedance 16 Connect the equipment as shown in Figure 6-19. Figure 6-19. Receiver R Input Impedance Test Calibration Test Setup 17 Press PRESET, MEAS, INPUT PORTS, A/B, MENU, NUMBER of POINTS, 1601, x1. 18 Press CAL, CALIBRATE MENU, S11 1-PORT. Repeat Step 4 through Step 6 to perform a S11 1-port calibration. 19 Connect the equipment as shown in Figure 6-20. Figure 6-20.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy 20 Repeat Step 7 through Step 12 to measure the receiver R input impedance. In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy Figure 6-21. Receiver Magnitude Dynamic Accuracy Required Equipment Measuring Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8902A Step Attenuator, 110 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8496A Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010 Cable (4), 50Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . .
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy c Press AVG, IF BW, 10, x1. 4 Connect the equipment as shown in Figure 6-22. Figure 6-22. Channel A Magnitude Dynamic Accuracy Test Setup 5 On the Agilent 8702D Option 011: a Press MEAS, INPUT PORTS, A, MENU, POWER. b Using the analyzer front panel knob, adjust the test port power for a reading of –30.000 dBm4 on the Agilent 8902A. c Wait for the reading to stabilize.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy NOTE Ignore the UNCAL displayed on the Agilent 8902A. This indicates that you are making an approximate power level measurement. However, this is not a concern because you will be making a relative power measurement. 6 On the Agilent 8902A, press the blue shift key and SET REF. 7 On the Agilent 8702D Option 011: a Press CAL, CALIBRATE MENU, RESPONSE, THRU. b After the analyzer finishes the measurement, press DONE: RESPONSE.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy 18 Connect the equipment as shown in Figure 6-23. Figure 6-23. Channel B Magnitude Dynamic Accuracy Test Setup 19 On the Agilent 8702D Option 011: a Press MENU, POWER. b Using the analyzer front panel knob, adjust the test port power for a reading of 0.000 dB on the Agilent 8902A. c Wait for the reading to stabilize. 20 On the Agilent 8902A, press the blue shift key and SET REF.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy a Press CAL, CALIBRATE MENU, RESPONSE, THRU. b After the analyzer finishes the measurement, press DONE: RESPONSE. c Press SAVE/RECALL, SAVE STATE. 22 Set the Agilent 8496A to 0 dB. 23 On the Agilent 8702D Option 011, press MENU, TRIGGER MENU, SINGLE. 24 Write the mean value in the "Receiver Channel B" column of the "Performance Test Record." 25 Write the Agilent 8902A readout in the "8902A" column of the "Performance Test Record.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy Figure 6-24. Channel R Magnitude Dynamic Accuracy Test Setup 31 Set the Agilent 8496A to 20 dB. 32 On the Agilent 8702D Option 011: a Press CONTINUOUS, MEAS, INPUT PORTS, R, MENU, POWER. b Using the analyzer front panel knob, adjust the test port power for a reading of 0.000 dB on the Agilent 8902A. c Wait for the reading to stabilize. 33 On the Agilent 8902A, press the blue shift key and SET REF.
Performance Tests – Option 011 12. Receiver Magnitude Dynamic Accuracy 36 On the Agilent 8702D Option 011, press MENU, TRIGGER MENU, SINGLE. 37 Write the mean value in the "Receiver Channel R" column of the "Performance Test Record." 38 Write the Agilent 8902A readout in the "8902A" column of the "Performance Test Record." 39 Calculate the difference between the "Receiver Channel R" and the "8902A".
Performance Tests – Option 011 13. Receiver Phase Dynamic Accuracy a Perform the "IF Amplifier Correction Constants" and "ADC Offset Correction Constants" procedures (located in the "Adjustments and Correction Constants" chapter of the Agilent 8753D Option 011 Service Guide). b Repeat this performance test. c If it still fails, replace the A10 Digital IF assembly. d Repeat the two adjustment procedures mentioned in this step and then repeat this performance test. 13.
Performance Tests – Option 011 14. Receiver Magnitude Compression CalculatedValue A = 10 --------------------------------------------- 20 4 Write the result in the "A" column of the "Performance Test Record." 5 For every receiver input power level, calculate "B" using the following formula: B=1–A 6 Write the result in the "B" column of the "Performance Test Record.
Performance Tests – Option 011 14. Receiver Magnitude Compression Additional Required Equipment for Analyzers with Option 006 Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A Procedure Channel A Magnitude Compression 1 Zero and calibrate the power meter. 2 Connect the equipment as shown in Figure 6-26.
Performance Tests – Option 011 14. Receiver Magnitude Compression Figure 6-26.
Performance Tests – Option 011 14. Receiver Magnitude Compression 3 Press PRESET, AVG, IF BW, 10, x1. 4 Press MENU, CW FREQ, 50, M/µ. Set the calibration factor on the power meter for this CW frequency. 5 Press POWER, –4, x1. Use the analyzer’s front panel knob to adjust the source power for a power meter reading of –20.00 dBm. Write the source power value, which appears on the analyzer’s display, in the “Start Power” column of the “Performance Test Record”. 6 Press 15, x1.
Performance Tests – Option 011 14. Receiver Magnitude Compression 20 Press STOP and enter the “Stop Power” value as determined in Step 18. 21 Press MENU, TRIGGER MENU, SINGLE. 22 At the end of the sweep, press SCALE REF, AUTO SCALE. 23 Press MARKER, ∆ MODE MENU, ∆REF=1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 24 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MIN.
Performance Tests – Option 011 14. Receiver Magnitude Compression Channel B Magnitude Compression 27 Connect the equipment as shown in Figure 6-27. Figure 6-27.
Performance Tests – Option 011 14. Receiver Magnitude Compression 28 Press MEAS, INPUT PORTS, B/R. 29 Repeat Step 26 for the CW frequencies listed in the “Performance Test Record”. Channel R Magnitude Compression 30 Connect the equipment as shown in Figure 6-28.
Performance Tests – Option 011 14. Receiver Magnitude Compression Figure 6-28. Channel R Magnitude Compression Test Setup 31 Repeat Step 26 for the CW frequencies listed in the “Performance Test Record”.
Performance Tests – Option 011 15. Receiver Phase Compression In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests – Option 011 15. Receiver Phase Compression Procedure Channel A Phase Compression 1 Zero and calibrate the power meter. 2 Connect the equipment as shown in Figure 6-29. Figure 6-29.
Performance Tests – Option 011 15. Receiver Phase Compression 3 Press PRESET, AVG, IF BW, 10, x1, FORMAT, PHASE. 4 Press MENU, CW FREQ, 50, M/µ. Set the calibration factor on the power meter for this CW frequency. 5 Press POWER, –4, x1. 6 Use the analyzer’s front panel knob to adjust the source power for a power meter reading of –20.00 dBm. Write the source power value, which appears on the analyzer’s display, in the “Start Power” column of the “Performance Test Record”. 7 Press 15, x1.
Performance Tests – Option 011 15. Receiver Phase Compression keypad to enter the “Start Power” value as determined in Step 18. 21 Press STOP and enter the “Stop Power” value as determined in Step 19. 22 Press MENU, TRIGGER MENU, SINGLE. 23 At the end of the sweep, press SCALE REF, AUTO SCALE. 24 Press MARKER, ∆ MODE MENU, ∆REF=1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 25 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MIN.
Performance Tests – Option 011 15. Receiver Phase Compression Channel B Phase Compression 28 Connect the equipment as shown in Figure 6-30. Figure 6-30. Channel B Phase Compression Test Setup 29 Press MEAS, INPUT PORTS, B/R. 30 Repeat Step 27 for the other CW frequencies listed in the “Performance Test Record.
Performance Tests – Option 011 15. Receiver Phase Compression Channel R Phase Compression 31 Connect the equipment as shown in Figure 6-31. Figure 6-31. Channel R Phase Compression Test Setup 32 Repeat Step 27 for the other CW frequencies listed in the “Performance Test Record.
Performance Tests – Option 011 16. Source and Receiver Harmonics (Option 002 Only) In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
Performance Tests – Option 011 16. Source and Receiver Harmonics (Option 002 Only) Figure 6-32. Source Harmonics Test Setup 2 Press PRESET, MENU, POWER, 20, x1. 3 Press START, 16, M/µ, STOP, 1.5, G/n to set the frequency range for searching the source 2nd harmonic. 4 Only for Analyzers with Option 006: Press STOP, 3, G/n. 5 Press AVG, IF BW, 10, x1 to set the IF bandwidth to 10 Hz. 6 Press MEAS, INPUT PORTS, A.
Performance Tests – Option 011 16. Source and Receiver Harmonics (Option 002 Only) 13 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, DISPLAY. 14 After one sweep, press DATA→MEMORY, DATA/MEM to normalize the trace and record the fundamental. 15 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD to prepare the analyzer for making a 3rd harmonic measurement. 16 After one sweep, press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 17 Write the marker 1 value on the "Performance Test Record.
Performance Tests – Option 011 16. Source and Receiver Harmonics (Option 002 Only) 24 Move the receiver A input cable to receiver input B. 25 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, STOP, 1.5, G/n. 26 Only for Analyzers with Option 006: Press STOP, 3, G/n. 27 Repeat Step 6 through Step 9 to measure the source and receiver worst case 2nd harmonic for the receiver input B. Press MEAS, INPUT PORTS, B in Step 6.
Performance Tests – Option 011 17. Receiver Magnitude Frequency Response (Option 002 Only) for the receiver input A. Press MEAS, INPUT PORTS, A in Step 6. 36 Repeat Step 11 through Step 16 to measure the receiver worst case 3nd harmonic for the receiver input A. In Case of Difficulty WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
Performance Tests – Option 011 17. Receiver Magnitude Frequency Response (Option 002 Only) Additional Equipment Required for Option 006 Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A Procedure Source Harmonics 1 Connect the equipment as shown in Figure 6-35. Figure 6-35. Magnitude Frequency Response Test Setup 2 Zero and calibrate the power meter. Set the power meter to measure dBm.
Performance Tests – Option 011 17. Receiver Magnitude Frequency Response (Option 002 Only) Test Record.” 8 Press B. Write the marker 1 reading, which appears on the analyzer display, in the “Input B Value” column on the “Performance Test Record.” 9 On the external source, press 48, MHz. Record the power meter reading on the “Performance Test Record.” 10 On the analyzer, press SYSTEM, HARMONIC MEAS, THIRD. Write the marker 1 reading in the “Input B Value” column on the “Performance Test Record.
Performance Tests – Option 011 Performance Test Records - Option 011 Performance Test Records - Option 011 The following tables are to record the performance test records for each of the Agilent 8702D Option 011 tests. Test 1. Source Frequency Range and Accuracy CW Frequency (MHz) Lower Limit (MHz) Measured Value (MHz) Upper Limit (MHz) Measurement Uncertainty (MHz) 0.3 0.299 997 ____________________ 0.300 003 ± 0.000 000 360 5.0 4.999 950 ____________________ 5.000 050 ± 0.000 006 16.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 2.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 2. Source Power Range, Linearity, and Accuracy (Power Range and Power Linearity) (2 of 3) Source Power Level (dBm) Power Offset (dB) Path Loss (dB) Measured Value (dB) Power Linearity (dB) Specification (dB) Meas Uncertainty (dB) +1 +9 ________ ________ ________ ± 0.25 ± 0.15 +3 +7 ________ ________ ________ ± 0.25 ± 0.15 +5 +5 ________ ________ ________ ± 0.25 ± 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 2. Source Power Range, Linearity, and Accuracy (Power Range and Power Linearity) (3 of 3) Source Power Level (dBm) + 20 Power Offset (dB) Path Loss (dB) Measured Value (dB) Power Linearity (dB) Specification (dB) Meas Uncertainty (dB) – 10 ________ ________ ________ ± 0.5 ± 0.15 + 13 –3 ________ ________ ________ ± 0.5 ± 0.15 + 18 –8 ________ ________ ________ ± 0.5 ± 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 2. Source Power, Range, Linearity and Accuracy (Power Level Accuracy) Path Loss (dB) Calibrated Power Level (dB) Measured Value (dB) Power Level Accuracy (dB) 0.300 _______ _______ _______ _______ ± 1.0 ± 0.15 20.000 _______ _______ _______ _______ ± 1.0 ± 0.10 50.000 _______ _______ _______ _______ ± 1.0 ± 0.10 100.000 _______ _______ _______ _______ ± 1.0 ± 0.10 200.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 3. Receiver Minimum R Channel Level CW Frequency Specification (dB) Marker Value (dB) Measurement Uncertainty (dB) 300 kHz < –35 ____________________ ± 1.0 3.29 MHz < –35 ____________________ ± 1.0 3.31 MHz < –35 ____________________ ± 1.0 15.90 MHz < –35 ____________________ ± 1.0 16.10 MHz < –35 ____________________ ± 1.0 30.90 MHz < –35 ____________________ ± 1.0 31.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 4. Receiver Minimum R Channel Level for External Source Mode CW Frequency (MHz) Frac-N VCO Lower Limit (MHz) Frac-N VCO Upper Limit (MHz) Measured Value (MHz) Measurement Uncertainty 10 49.496 50.496 ____________ N/A 20 37.620 38.380 ____________ N/A 100 49.005 49.995 ____________ N/A 1 000 36.630 37.370 ____________ N/A 3 000 58.216 59.392 ____________ N/A 4 000 39.198 39.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 5. Receiver Channel Noise Floor Level Frequency Range IF Bandwidth Specification (dBm) Calculated Value Measurement Uncertainty Receiver Channel A 300 kHz – 3.0 GHz 3 kHz – 90 ____________ N/A 300 kHz – 3.0 GHz 10 Hz – 110 ____________ N/A 300 kHz – 3.0 GHz 10 Hz – 110 ____________ N/A 300 kHz – 3.0 GHz 3 kHz – 90 ____________ N/A 50 kHz – 3.0 GHz 3 kHz – 90 ____________ N/A 50 kHz – 3.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 7. Receiver Magnitude Frequency Response CW Frequency Power Meter Reading Example – 10.0 – 10.14 – 10.09 – 10.10 0.14 ±1 ± 0.05 300 kHz _______ _______ _______ _______ _______ ±1 ± 0.05 5 MHz _______ _______ _______ _______ _______ ±1 ± 0.05 16 MHz _______ _______ _______ _______ _______ ±1 ± 0.05 31 MHz _______ _______ _______ _______ _______ ±1 ± 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 8. Receiver Phase Frequency Response Frequency Range Ratio Specification Measured Value Measurement Uncertainty 300 kHz – 3 GHz A/R ± 3° _____________ ± 0.35° 300 kHz – 3 GHz B/R ± 3° _____________ ± 0.35° 300 kHz – 3 GHz A/B ± 3° _____________ ± 0.35° 3 GHz – 6 GHz A/B ± 10° _____________ ± 0.35° 3 GHz – 6 GHz A/R ± 10° _____________ ± 0.35° 3 GHz – 6 GHz B/R ± 10° _____________ ± 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 9. Receiver Input Crosstalk (2 of 2) Frequency Range Specification (dB) Marker Value Measurement Uncertainty Option 006 R into A Crosstalk 3.0 GHz – 4.5 GHz – 82 _______________ N/A 4.5 GHz – 6.0 GHz – 75 _______________ N/A 3.0 GHz – 4.5 GHz – 82 _______________ N/A 4.5 GHz – 6.0 GHz – 75 _______________ N/A 3.0 GHz – 4.5 GHz – 82 _______________ N/A 4.5 GHz – 6.0 GHz – 75 _______________ N/A 3.0 GHz – 4.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 10. Receiver Trace Noise CW Frequency (GHz) Ratio Specification (rms) Measured Value Measurement Uncertainty 3 A/R ____________________ ≤ 0.006 dB N/A 3 B/R ____________________ ≤ 0.006 dB N/A 3 A/B ____________________ ≤ 0.006 dB N/A 3 A/B ____________________ ≤ 0.038° N/A 3 B/R ____________________ ≤ 0.038° N/A 3 A/R ____________________ ≤ 0.038° N/A Option 006 6 A/R ____________________ ≤ 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 12. Receiver Magnitude Dynamic Accuracy (1 of 2) Receiver Input Power (dB) Agilent 8496A Setting (dB) Receiver Channel A (dB) Agilent 8902A (dB) Magnitude Value (dB) Spec. (dB) Meas. Uncer. (dB) Channel A – 10 0 _________ _________ _________ ≤ 0.075 ± 0.010 – 20 10 _________ _________ _________ ≤ 0.059 ± 0.020 – 30 20 _________ _________ _________ ≤ 0.050 ± 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 12. Receiver Magnitude Dynamic Accuracy (2 of 2) Receiver Input Power (dB) Agilent 8496A Setting (dB) Receiver Channel A (dB) Agilent 8902A (dB) Magnitude Value (dB) Spec. (dB) Meas. Uncer. (dB) – 90 80 _________ _________ _________ ≤ 0.680 ± 0.180 – 100 90 _________ _________ _________ ≤ 1.950 ± 0.200 – 110 100 _________ _________ _________ ≤ 5.200 ± 0.280 – 10 0 _________ _________ _________ ≤ 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 13. Receiver Phase Dynamic Accuracy (2 of 2) Receiver Input Power (dB) Magnitude Value (dB) A B Calculated Value (degrees) Spec. (degrees) Mea. Uncer. – 100 _________ _________ _________ _________ ≤ 16.00 N/A – 110 _________ _________ _________ _________ ≤ 51.50 N/A – 10 _________ _________ _________ _________ ≤ 0.80 N/A – 20 _________ _________ _________ _________ ≤ 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 14. Receiver Magnitude Compression (1 of 2) CW Frequency Start Power (dBm) Stop Power (dBm) Measured Value (dB) Specification (dB) Measurement Uncertainty Channel A 50 MHz _________ _________ _________ ≤ 0.32 N/A 1 GHz _________ _________ _________ ≤ 0.32 N/A 2 GHz _________ _________ _________ ≤ 0.32 N/A 3 GHz _________ _________ _________ ≤ 0.32 N/A 50 MHz _________ _________ _________ ≤ 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 14. Receiver Magnitude Compression (2 of 2) CW Frequency Start Power (dBm) Stop Power (dBm) Measured Value (dB) Specification (dB) Measurement Uncertainty 5 GHz _________ _________ _________ ≤ 0.32 N/A 6 GHz _________ _________ _________ ≤ 0.32 N/A 4 GHz _________ _________ _________ ≤ 0.32 N/A 5 GHz _________ _________ _________ ≤ 0.32 N/A 6 GHz _________ _________ _________ ≤ 0.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 15. Receiver Phase Compression (2 of 2) CW Frequency Start Power (dBm) Stop Power (dBm) Measured Value (degrees) Specification (degrees) Measurement Uncertainty 1 GHz _________ _________ _________ ≤ 5.2 N/A 2 GHz _________ _________ _________ ≤ 5.2 N/A 3 GHz _________ _________ _________ ≤ 5.2 N/A 4 GHz _________ _________ _________ ≤ 5.2 N/A 5 GHz _________ _________ _________ ≤ 5.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 16. Source and Receiver Harmonics (Option 002 Only) (1 of 2) Test Description Harmonic Specification (dBc) Measurement Value Measurement Uncertainty (dB) Source Harmonics 1.5 2nd < – 25 _______________ ±1 1.0 3rd < – 25 _______________ ±1 1.5 A: 2nd < – 15 _______________ ±1 1.0 A: 3rd < – 30 _______________ ±1 1.5 B: 2nd < – 15 _______________ ±1 1.0 B: 3rd < – 30 _______________ ±1 1.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 16.
Performance Tests – Option 011 Performance Test Records - Option 011 Test 17. Receiver Magnitude Response (Option 002 Only) Analyzer Frequency Ext. Source Frequency Power Meter Value Input A Value Input B Value Max. Diff. Spec Spec. (dB) Meas. Uncer. (dB) 16 MHz 32 MHz _______ _______ _______ _______ ±1 ± 0.05 16 MHz 48 MHz _______ _______ _______ _______ ±1 ± 0.05 31 MHz 62 MHz _______ _______ _______ _______ ±1 ± 0.
7 If the display does not light 7-2 If the fan does not run 7-3 If data entry keys don’t respond 7-3 If there is no RF signal 7-4 Returning the Instrument for Service 7-5 Agilent Technologies Service Offices 7-8 If You Encounter a Problem
If You Encounter a Problem If You Encounter a Problem If You Encounter a Problem If you have difficulty when installing or using the Agilent 8702D, check the following list of commonly encountered problems and troubleshooting procedures. Contact the nearest Agilent Technologies office for service, if necessary. A list of Agilent Technologies service offices is provided on page 7-8.
If You Encounter a Problem If the fan does not run WARNING The power cord is connected to internal capacitors that may remain live for 10 seconds after you disconnect the cord from the power supply. Figure 7-1. Line Fuse Removal and Replacement WARNING For continued protection against fire hazard, replace the fuse with the same type and rating. If the fan does not run 1 Switch the LINE power switch to the off position. 2 Check that the fan blades are not jammed or obstructed.
If You Encounter a Problem If there is no RF signal The “entry off” is enabled after you press the ENTRY OFF key. To return to normal entry mode, press any function key that has a numeric parameter associated with it. For example, START. ❒ Check that none of the keys are mechanically stuck. ❒ Check that the selected key is designed to accept data. For example, SCALE REF accepts data, but SYSTEM does not. ❒ Ensure that you are not attempting to enter a value that is greater than or less than its limit.
If You Encounter a Problem Returning the Instrument for Service Returning the Instrument for Service The instructions in this section show you how to properly return the instrument for repair or calibration. Always call the Agilent Technologies Instrument Support Center first to initiate service before returning your instrument to a service office. This ensures that the repair (or calibration) can be properly tracked and that your instrument will be returned to you as quickly as possible.
If You Encounter a Problem Returning the Instrument for Service information should be returned with the instrument. • Type of service required. • Date instrument was returned for repair. • Description of the problem: • Whether problem is constant or intermittent. • Whether instrument is temperature-sensitive. • Whether instrument is vibration-sensitive. • Instrument settings required to reproduce the problem. • Performance data. • Company name and return address.
If You Encounter a Problem Returning the Instrument for Service Sealed Air Corporation (Commerce, California 90001). Air Cap looks like a plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap™ to reduce static electricity. Wrapping the instrument several times in this material will protect the instrument and prevent it from moving in the carton. 4 Seal the carton with strong nylon adhesive tape. 5 Mark the carton “FRAGILE, HANDLE WITH CARE”. 6 Retain copies of all shipping papers.
If You Encounter a Problem Agilent Technologies Service Offices Agilent Technologies Service Offices Before returning an instrument for service, call the Agilent Technologies Instrument Support Center at (800) 403-0801, visit the Test and Measurement Web Sites by Country page at http://www.tm.agilent.com/tmo/country/English/ index.html, or call one of the numbers listed below. Agilent Technologies Service Numbers Austria 01/25125-7171 Belgium 32-2-778.37.
Index A accuracy and range of frequency, 4-3 accuracy and range of frequency (Option 011), 6-3 accuracy of power test, 4-6 Agilent offices, 7-8 B backup disk, 1-26 C cabinet, cleaning, iv, 1-2 calibration coeficients test, 4-34 care of cabinet, iv, 1-2 checking the operation, 1-16 classification product, iv cleaning cabinet, iv, 1-2 clock, setting, 1-15 connecting a plotter, 1-12, 1-14 connecting a printer, 1-12, 1-13 copying the EEPROM disk, 1-26 D damaged shipment, 1-3 disk, backup, 1-26 6-21 test po
Index test port input frequency response, 4-19, 6-21 test port input impedance, 4-44, 6-35 test port input noise floor level, 4-15 test port output frequency range and accuracy, 4-3 test port output power accuracy, 4-6 test port output power range and linearity, 4-8, 6-5 test port receiver magnitude compression, 4-57, 6-48 test port receiver magnitude dynamic accuracy, 4-48, 6-39 test port receiver phase compression, 4-59, 6-56 test port receiver phase dynamic accuracy, 4-55, 6-47 I IEC Publication 61010-
Index printer type, 1-13 problems, data entry controls do not respond, 7-3 problems, power-up display lights, but fan does not start, 1-7, 7-3 problems, RF output no RF signal at either PORT 1 or PORT 2, 7-4 procedures calibration coefficients, 4-34 external source mode frequency range, 4-5 minimum R channel level, 4-10, 6-9, 6-12 source frequency range and accuracy, 6-3 system trace noise (with option 006), 4-42 system trace noise (without option 006), 4-40, 6-33 test port crosstalk, 4-29, 6-26 test port
Index test port output power range and linearity, 4-8, 6-5 test port receiver magnitude compression, 4-57, 6-48 test port receiver magnitude dynamic accuracy, 4-48, 6-39 test port receiver phase compression, 4-59, 6-56 test port receiver phase dynamic accuracy, 4-55, 6-47 system trace noise test (with option 006), 4-42 system trace noise test (without option 006), 4-40, 6-33 T test port crosstalk test, 4-29, 6-26 test port input impedance test, 4-44, 6-35 test port output frequency range and accuracy test