user manual
Chapter 3: Testing Performance
To test time measurement accuracy
3–20
26
Clear measurement statistics as in step 18 and restart the measurement.
27 The delta time reading should be 1 µs ± 2.595 ns, minimum 997.4 ns and maximum 
1.0026 
µs. Record the minimum and maximum readings in the Performance Test 
Record.
28 Select Measurement Definitions from the Customize submenu of the Measure menu. 
(If you have code Revision A.03.00 or higher, select Delta Time from the Measurements 
Definitions submenu of the Measure menu.)
29 Set the To Edge # to 6 with Direction Rising and Threshold Middle. Click Close.
See figure 3-13.
Figure 3-13
New Measurement Settings for Delta Time Measurement
30 Clear measurement statistics as in step 18 and restart the measurement.
31 The delta time readings should be 5 µs ± 2.875 ns, minimum 4.9971 µs and maximum 
5.0029 
µs. Record the minimum and maximum readings in the Performance Test 
Record.