User`s guide
Quick Reference 1
Series N6700 User’s Guide 9
Output a
utoranging
capa
bility
Available on Agilent N676xA and N675xA power modules.
Autoranging supplies the maximum rated power over a continuous range of
voltage and current settings.
Output On/Off sequencing
A turn
-on/turn-off delay capability for each output allows output on/off
sequencing.
Remote voltage sensing
Two remote sensing terminals are provided for each output. When shipped,
the remote sense jumpers are included in a separate bag. See Chapter 2.
Output
protection
Each output has over
-voltage, over-current, and over-temperature
protection. Over-voltage and over-current protection are programmable.
Multiple
-Quadrant
operation
Available on Agilent N678xA
SMU and N6783A power modules.
2- and 4-quadrant operation provides source and sink output capability.
Measurement Features
Multiple
-output/Single-
output meter display
Switch between a 4
-output summary view and a 1-output detailed view of
power supply information. All power modules display real
-time output
voltage and current measurements as well as status information.
Seamless measurement
autoranging
Available on Agilent N6781A and N6782A
SMU power modules.
Output
measurements seamlessly autorange between ranges – however,
the 10 μA current range must be selected manually.
Microampere
current
measurements
Available on Agilent
N6761A, N6762A, and N678xA SMU power modules.
Current measurement can be made as low as 1 μA in the 10 μA range.
Fast d
igitizing
Available on Agilent N678xA
SMU power modules.
5.12 μs/sample for one parameter; 10.24 μs/sample for two parameters.
System Features
Choice of three interfaces
GPIB (IEEE-488), LAN, and USB remote programming interfaces are built in.
Built
-in Web server
A built
-in Web server lets you control the instrument directly from an
internet browser on your computer.
SCPI language
The instrument is compatible with the Standard Commands for
Programmable Instruments (SCPI).
Front panel I/O setup
Menus let you set up GPIB and LAN parameters from the front panel.
Real
-time status
information
The front panel indicates the status of each output. It also indicates when a
protection shut-down has occurred.
Module identification
Each module has identifying data stored in non
-volatile memory.
Information includes model number, serial number, and options.
Universal AC
input
Mainframes have
universal input voltage capability
with active power factor
correction.