User`s guide
Operating the Power System 4
Series N6700 User’s Guide 71
You can test the TRAN_active bit in the operation status register to
know when the transient trigger system has returned to the idle
state.
Front Panel: SCPI Command:
Not Available To query the TRAN_active bit (bit 6):
STAT:OPER:COND?(@1)
If a bit value of 64 is returned in the query, the TRAN_active bit is
true, and the transient action is NOT complete. When the
TRAN_active bit is false, the transient action is complete. Refer to the
N6700 Programmer’s Reference Help file for more information.
Generate Trigger Out Signals
Each output channel can generate trigger signals that can be used by
other output channels, or routed to a pin on the digital port that has
been configured as a trigger output (TOUT). Use the following
commands to program transient trigger signals that are generated
when an output Step occurs:
Front Panel: SCPI Command:
Use the Channel key to select the
channel that is the trigger source.
Select Transient\Step.
Check Enable Trigger Output. Then
press Select.
To program channel 3’s step function
to generate a trigger signal, use
STEP:TOUT ON,(@3)