User`s guide
4 Operating the Power System
66 Series N6700 User’s Guide
Front Panel: SCPI Command:
Not Available To leave the relays closed when output 1 is turned off:
OUTP OFF,NOR,(@1)
To leave the relays open when output 1 is turned on:
OUTP ON,NOR,(@1)
If you have Option 760 installed, you can also reverse the polarity of
the output and sense terminals. Note that this command briefly turns
the output off while the output and sense terminal polarities are
switched. Also note that if this option is installed in Model N6742B,
the maximum output current will be limited to 10A.
Front Panel: SCPI Command:
Select Output\Advanced\Pol.
Then check the Reverse box.
Uncheck the Reverse box to
return the polarity to normal.
To switch the output and sense terminal
polarities of output 1:
OUTP:REL:POL REV,(@1)
To return the polarities back to normal:
OUTP:REL:POL NORM,(@1)
NOTE
When the output polarity has been reversed, the voltage meter on the front
panel display will show a bar over the voltage reading:
Set the Agilent N6781A Output Resistance
NOTE
This information applies to Agilent Model N6781A only.
Output resistance programming is primarily used in battery testing
applications, and only applies in Voltage priority mode. Values are
programmed in Ohms, from – 40 mΩ to + 1 Ω.
Front Panel: SCPI Command:
Select
Output\Advanced\Resistance.
Enter a value and press Select.
To set the output resistance to 0.5
Ω:
OUTP:RES 0.5,(@1)
Set the Output Bandwidth
NOTE
This information applies to Agilent Models N678xA SMU only.
Agilent Models N678xA SMU have several voltage bandwidth modes
to allow you to optimize output response time with capacitive loads.
The Low bandwidth setting provides stability with a wide range of
output capacitors. Additional bandwidth modes provide faster output
response when the output capacitance is restricted to smaller values.
10.001V