User`s guide

Installation 2
Series N6700 User’s Guide 41
Setting the Outputs
Outputs connected together in series cannot be grouped.
To program outputs connected in series, first program the current
limit of each output to the total desired current limit point. Then
program the voltage of each output so that the sum of both voltages
equals the total desired operating voltage. The simplest way to
accomplish this is to program each output to one half of the total
desired operating voltage.
NOTE
The operating mode of each output is determined by the output’s programmed
settings, operating point, and load condition. Because these conditions may
change during series operation, the front panel status indicator will reflect
these changes. This is normal. Momentary status changes are also normal.
Effect on Specifications
Specifications for outputs operating in series can be obtained from
the specifications for single outputs. Most specifications are
expressed as a constant or a percentage (or ppm) plus a constant.
For series operation, the percentage portion remains unchanged
while constant portions or any constants are changed as indicated.
Voltage
All series specifications referring to voltage are twice the single output
specification except for programming resolution, which is the same as for a
single output.
Current All series specifications referring to current are the same as for a single
output except for CC load effect, CC load cross regulation, CC source effect,
and CC short term drift. These are twice the current programming accuracy
(including the percentage portion) at all operating points.
Load Transient
Recovery Time
Load transient specifications are typically twice the single output.
WITH LOCAL SENSING WITH REMOTE SENSING
SENSE
JUMPERS
INSTALLED
LOAD
OUTPUT 1OUTPUT 3
LOAD
+
SENSE
JUMPER
INSTALLED
TWIST LEADS
+
TWIST LEADS
+
+
+S +LS -LS -S
OUTPUT 1OUTPUT 3
+
+
+S +LS -LS -S
+S +LS -LS -S
+S +LS -LS -S