Specifications

Protection Features
Each N6700 module is protected
against over-voltage, over-
current, and over-temperature.
A fault condition in one
module can be detected within
10 microseconds by other mod-
ules so that they can be quickly
shut down to avoid hazardous
conditions on your DUT.
Connectivity
The N6700 MPS comes standard
with GPIB, USB 2.0, and
10/100 Base-T Ethernet LAN
interfaces. While GPIB is best
suited for use with existing
systems, Agilent offers USB
and LAN to allow you to take
advantage of the availability,
speed, and ease-of-use of
common computer industry
standard interfaces.
Security
When used in systems running
GPIB, the LAN and/or USB
interfaces can be disabled
for extra security. Also, all
non-volatile RAM data and
settings can be cleared from
the front panel.
Control from any Browser
The N6700 can be controlled
via a standard web browser.
The N6700 contains a web
server that provides web pages
for monitor, control, and setup
of the MPS.
Output Sequencing
Each DC power module can be
individually set to turn on or
to turn off with a delay. By
adjusting the delay times and
then commanding the N6700 to
turn on, you can set the N6700
modules to sequence on in a
particular order. The same
sequencing capability is avail-
able to shut down the modules
in a particular order.
3
Module 1
Output on
V1
V2
V3
Module 2
Delay 2
Delay 3
Module 3
Figure 2. Output Sequencing