Specifications
18
N6700B MPS Mainframe
Output Response
Characteristics Command processing time ≤ 1 ms from receipt of command to start of the output change
Protection Response
Characteristics Inhibit Input 5 µs (from receipt of inhibit to start of shutdown)
Fault on coupled outputs < 10 µs (from receipt of fault to start of shutdown)
Digital Control
Characteristics Maximum voltage ratings +16.5 VDC/-5 VDC between pins (pin 4 is internally connected to chassis ground).
Pins 1 and 2 as FLT output Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 0.14 mA @ 16.5 VDC
Pins 1 thru 8 as digital/trigger outputs Maximum low-level output voltage = 0.5 V @ 4 mA; 1 V @ 50 mA; 1.75 V @ 100 mA
(pin 4 = common) Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Pins 1 thru 8 as digital/trigger inputs Maximum low-level input voltage = 0.8 V
and pin 3 as INH input Minimum high-level input voltage = 2 V
(pin 4 = common) Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Interface Capabilities
GPIB SCPI - 1993, IEEE 488.2 compliant interface
USB 2.0 Requires Agilent I/O Library version M.01.01.04
10/100 LAN Requires Agilent I/O Library version L.01.01
Web server Built-in Web server - requires Internet Explorer 5+ or Netscape 4.x
Environmental Conditions
Operating environment Indoor use, installation category II
1
, pollution degree 2
Temperature range 0
˚
C to 55
˚
C (output derated above 40
˚
C)
Relative humidity Up to 95%
Altitude Up to 2000 meters
Storage temperature -30
˚
C to 70
˚
C
1
Category II for AC input.
Supplemental Characteristics (Continued)
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or type testing.
All characteristics are typical unless otherwise noted.