Specifications
13
Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C, and derated above 40°C.
N6751A / N6752A N6761A / N6762A
DC Output Ratings
Voltage 50 V 50 V
Current 5 A / 10 A 1.5 A / 3 A
Power 50 W / 100 W 50 W / 100 W
Output Ripple
and Noise (PARD)
(from 20 Hz – 20 MHz) CV peak-to-peak
1
6 mV 6 mV
CV rms 1 mV 1 mV
Load Effect
(Regulation) Voltage 2 mV 0.5 mV
(for any output load change, Current 2 mA 30 µA
with a maximum load-lead
drop of 1 V per lead)
Source Effect
(Regulation) Voltage 1 mV 1 mV
Current 1 mA 30 µA
Programming
Accuracy
(at 23°C ±5°C Voltage high range 0.06% + 19 mV 0.016% + 6 mV
after 30 minute Voltage low range (≤ 5.5 V) N/A 0.015% + 1.5 mV
warm-up. Applies Current high range 0.1% + 20 mA 0.04% + 200 µA
from min. to max. Current low range (≤ 100 mA, @ 0 - 7 V)
NOTE 2
N/A 0.04% + 15 µA
programming range) (≤ 100 mA, @ 0 - 50 V) N/A 0.04% + 55 µA
Measurement
Accuracy
(at 23°C ±5°C) Voltage high range 0.05% + 20 mV 0.016% + 6 mV
Voltage low range (≤ 5.5 V) N/A 0.016% + 1.5 mV
Current high range 0.1% + 4 mA 0.03% + 200 µA
Current low range (≤ 100 mA, @ 0 - 7 V)
NOTE 2
N/A 0.03% + 15 µA
NOTE 3
(≤ 100 mA, @ 0 - 50 V) N/A 0.03% + 55 µA
Load Transient
Recovery Time
(time to recover to within the settling band following a load change)
• from 60% to 100% and from 100% to 60% of full load for models N6751A & N6761A
• from 50% to 100% and from 100% to 50% of full load for models N6752A & N6762A.
Voltage settling band ± 75 mV ± 75 mV
Time < 100 µs < 150 µs
1
For typical values, refer to Supplemental Characteristics.
2
Requires a 3 minute wait if you have entered the 0- 7 V range from a voltage greater than 7 V..
3
Applies when measuring 4006 data points (SENSe:SWEep:POINts = 4096).