Specifications
Series N6700 Service Guide 9
2
Performance Verification
Verification Description .................................................................................... 10
Verification Procedure ...................................................................................... 13
Test Record Forms ............................................................................................ 22
The verification procedures described in this chapter verify that the
Agilent N6700 Modular Power System is operating normally and is
within published specifications.
NOTE
Perform the verification tests before calibrating your power system. If the power
system passes the verification tests, the unit is operating within its calibration
limits and does not need to be re-calibrated.
Verification procedures perform two primary functions:
Performance Tests
These tests verify that the power system is operating normally and meets the
specifications published in the Agilent N6700 Modular Power System Family
Specifications Guide.
Calibration Tests These tests verify that the power system is operating within its calibration
limits.
If the instrument fails any of the tests or if abnormal test results are
obtained, try calibrating the unit. If calibration is unsuccessful,
return the unit to an Agilent Technologies Service Center.