Specifications

4 Disassembly
84 Series N6700 Service Guide
Step 3. To install a module, align the module over the alignment pins, and
push it down onto the mainframe connector.
Step 4. Fasten the module to the mainframe. Install the two screws from the
power module or filler module at either end of the module. Because
the RFI strips are applying upward pressure, continue pushing down
on the module until the screws are fully tightened.
Step 5. Replace the blower cover when finished.
Accessing the Calibration switch
Step 1. Remove the blower cover. Remove three screws from the top of the
cover and two screws on the sides. Tilt the cover up and slide it out.
Step 2. Remove the interface cover. Remove the three screws along the left
side of the unit. Then remove the screw at the front of the cover. Lift
off the cover.
Step 3. The calibration switch is on the interface board near the ribbon
cable. To change the calibration switch settings, use a small
screwdriver to move the switches. Refer to chapter 3 for settings
information.
CAUTION
Do not use a pencil to move the switches. Any graphite dust that gets on the
switches will conduct electricity.
Step 4. Replace all covers when finished.
1 2 3 4
ON