Specifications
Calibration 3
Series N6700 Service Guide 79
Change the Admin/Calibration Password
The password must be numeric, and can be up to 15 digits in length.
After it has been changed, you can only enter the calibration mode by
providing the correct password.
If the password is lost or forgotten, access can be restored by setting
an internal switch that resets the password to zero (refer to
“Accessing the Calibration switch” in chapter 4). Once you have set
the password to zero with the switch, you can enter a new password
using the Admin menu or using the CAL:PASS command. The new
password will become active when you log out of the Admin Menu or
send CAL STAT OFF. After you have returned the switches to the
normal position, you can access the Admin menu with the new
password.
Front Panel: SCPI Command:
Select System\Admin\Login.
Enter the original password in the
Password field and press Select.
Select System\Admin\Password.
Enter a new password in the
Password
field.
Logout of the Admin menu to activate
the password.
Enter calibration mode using the
original password
CAL:STAT ON, <password>
To change the password:
CAL:PASS <NRf>
To exit calibration mode and
activate the password:
CAL:STAT OFF
NOTE
If the message “Locked out by internal switch setting” or “Calibration is
inhibited by switch setting” appears, the internal switch is set to prevent the
password from being changed (refer to “Accessing the Calibration switch” in
chapter 4).