Specifications

3 Calibration
70 Series N6700 Service Guide
Step 5. Select the second current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field should indicate:
“Enter P2 measured data”
CAL:LEV P2
*OPC?
Step 6. Measure the current with the Agilent 3458A and enter the value. This
value should be negative.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
1 mA Measurement Range
Step 1. Connect the Agilent 3458A in series with a 10 kΩ resistor to the
output (see equipment list in chapter 2 and Cal Setup D).
Step 2. Select the 1 mA range of output 1.
Front Panel:
SCPI Command:
Select
System\Admin\Cal\Curr\Imeas
Then select the Low2 range.
CAL:CURR:MEAS 0.001, (@1)
Step 3. Select the first current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field should indicate:
“Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 4. Measure the current with the Agilent 3458A and enter the value.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>
Step 5. Select the second current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field should indicate:
“Enter P2 measured data”
CAL:LEV P2
*OPC?
Step 6. Measure the current with the Agilent 3458A and enter the value. This
value should be negative.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>