Specifications

Calibration 3
Series N6700 Service Guide 69
Step 6. Measure the current with the Agilent 3458A and enter the value.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>
Step 7. Select the third current calibration point for the negative range.
Front Panel:
SCPI Command:
Select Next. The information field should indicate:
“Enter P3 measured data”
CAL:LEV P3
*OPC?
Step 8. Measure the current with the Agilent 3458A and enter the value. This
value should be negative.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Step 9. Repeat steps 1 through 8 for the 10 mA programming range of model
N6784A. This time, select the Low2 range.
100 mA Measurement Range
Step 1. Connect only the current measurement terminals of the Agilent
3458A to the output terminals (see Cal Setup D without the resistor).
Step 2. Select the 100 mA range of output 1.
Front Panel:
SCPI Command:
Select System\Admin\Cal\Curr\Imeas
Then select the Low1 range.
CAL:CURR:MEAS
0.1
, (@1)
Step 3. Select the first current calibration point.
Front Panel: SCPI Command:
Select Next. The information field should indicate:
“Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 4. Measure the current with the Agilent 3458A and enter the value.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>