Specifications

3 Calibration
68 Series N6700 Service Guide
Step 7. Select the third voltage calibration point for the negative range.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: Enter P3 measured data”
CAL:LEV P3
*OPC?
Step 8. Calculate the shunt current (I=V/R) and enter the data. The value
should be negative.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Low Range Current Programming
Step 1. Connect only the current measurement terminals of the Agilent
3458A to the output terminals (see Cal Setup D without the resistor).
Step 2. Select the low-current programming range. The value to program a
range must be the maximum current of the range. Models N6781A
and N6782A have 300 mA range. Model N6784A has a 100 mA and a
10 mA range. This example selects the 300 mA programming range of
output 1. For model N6784A, select the 100 mA programming range.
Front Panel: SCPI Command:
Select System\Admin\Cal\Curr\Iprog
Then select the Low1 range.
CAL:CURR 0.3, (@1)
Step 3. Select the first current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field should indicate:
“Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 4. Measure the current with the Agilent 3458A and enter the value. This
value may be positive or negative.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>
Step 5. Select the second current calibration point.
Front Panel: SCPI Command:
Select Next. The information field should indicate:
“Enter P2 measured data”
CAL:LEV P2
*OPC?