Specifications

Calibration 3
Series N6700 Service Guide 65
Positive Voltage Limit
Step 1. Select the positive voltage limit range. The value to program a range
must be the maximum voltage of the range to be calibrated. This
example selects the 20 V positive limit range of output 1.
Front Panel: SCPI Command:
Select
System\Admin\Cal\Volt\Vlim
Then select the High range.
CAL:VOLT:LIM:POS 20
, (@1)
Step 2. Select the first voltage calibration point.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: Enter “Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 3. Measure the output voltage with the DMM and enter the data.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the
data from the external DMM. Press Select
when done.
CAL:DATA <data>
Step 4. Select the second voltage calibration point.
Front Panel:
SCPI Command:
Select Next. The information field should
indicate: “Enter P2 measured data”
CAL:LEV P2
*OPC?
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the
data from the external DMM. Press Select
when done. Press Next to finish calibration.
CAL:DATA <data>
Step 6. Repeat steps 1 through 5 for the 6 V positive limit range. This time,
select the Low range.