Specifications

Calibration 3
Series N6700 Service Guide 63
Step 4. Select the second voltage calibration point.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: Enter P2 measured data”
CAL:LEV P2
*OPC?
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Step 6 only applies to models N6781A and N6782A.
Step 6. Repeat steps 1 through 5 for the 600 mV range. This time, select the
Low2 range.
Steps 7 through 9 only apply to model N6784A.
Step 7. Select the third voltage calibration point for the negative range.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: Enter P3 measured data”
CAL:LEV P3
*OPC?
Step 8. Measure the output voltage with the DMM and enter the data. The
value should be negative.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Step 9. Repeat steps 1 through 5, 7 and 8 for the 600 mV range. This time,
select the Low2 range.
Low Range Voltage Measurement
Step 1. Select the low-voltage measurement range. The value entered must be
the maximum voltage of the range to be calibrated. This example
selects the 1 V range of output 1.
Front Panel: SCPI Command:
Select System\Admin\Cal\Volt\Vmeas.
Then select the Low1 range.
CAL:VOLT:MEAS 1
, (@1)