Specifications
3 Calibration
62 Series N6700 Service Guide
Step 6. Measure the output voltage with the DMM and enter the data.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Steps 7 and 8 only apply to model N6784A.
Step 7. Select the third voltage calibration point for the negative range.
Front Panel:
SCPI Command:
Select Next. The information field should
indicate: “Enter P3 measured data”
CAL:LEV P3
*OPC?
Step 8. Measure the output voltage with the DMM and enter the data. The
value should be negative.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Low Range Voltage Programming
Step 1. Select the low-voltage programming range. The value entered must be
the maximum voltage of the range to be calibrated. This example
selects the 6 V range of output 1.
Front Panel: SCPI Command:
Select System\Admin\Cal\Volt\Vprog.
Then select the Low1 range.
CAL:VOLT 6
, (@1)
Step 2. Select the first voltage calibration point.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: “Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 3. Measure the output voltage with the DMM and enter the data. This
value may be positive or negative.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>