Specifications

3 Calibration
54 Series N6700 Service Guide
Step 6. Measure the output voltage with the DMM and enter the data.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>
Low Range Voltage Programming
Step 1. Select the low-voltage programming range. The value entered must be
the maximum voltage of the range to be calibrated. This example
selects the 5 V range of output 1.
Front Panel: SCPI Command:
Select System\Admin\Cal\Volt\Vprog.
Then select the Low1 range.
CAL:VOLT 5, (@1)
Step 2. Select the first voltage calibration point.
Front Panel: SCPI Command:
Select Next. The information field should indicate:
“Enter P1 measured data”
CAL:LEV P1
*OPC?
Step 3. Measure the output voltage with the DMM and enter the data. This
value may be positive or negative.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
CAL:DATA <data>
Step 4. Select the second voltage calibration point.
Front Panel: SCPI Command:
Select Next. The information field should
indicate: Enter P2 measured data”
CAL:LEV P2
*OPC?
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel: SCPI Command:
Select the Measured Data field. Enter the data
from the external DMM. Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>