Specifications
3 Calibration
52 Series N6700 Service Guide
Step 4. Set the electronic load or connect the appropriate resistor for the
second calibration point.
Step 5. Select the second current calibration point.
Front Panel: SCPI Command:
Select Next. The calibration is automatic.
CAL:LEV P2
*OPC?
Downprogrammer Calibration
Step 1. Remove all loads from the output. This procedure is automatic and
takes a few seconds.
Front Panel:
SCPI Command:
Select System\Admin\Cal\Misc\Dprog.
Then select Next.
CAL:DPRog (@1)
*OPC?
Enter a Calibration Date
Front Panel: SCPI Command:
Select System\Admin\Cal\Date.
Enter the calibration date in the Date
field. If desired, you can enter
alphanumeric data in this field.
To enter a calibration date:
CAL:DATE “<date>”, (@1)
Save and Log out of Calibration Mode
CAUTION
Storing calibration constants overwrites the existing ones in non-volatile
memory. If you are not sure you want to permanently store the new constants,
do not Save the data when you exit the calibration mode. The calibration will
then remain unchanged.
Front Panel: SCPI Command:
Select System\Admin\Cal\Save. Select Save
to save all calibration data.
Select System\Admin\Logout to exit
calibration mode.
To save calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF