Specifications

Calibration 3
Series N6700 Service Guide 45
Calibration Set-up
The following figures show the calibration test set-ups. Connect all
leads to the output terminals as shown in the following figure. Twist
the + and - load leads to reduce noise and inductance.
NOTE
Load lead lengths for models N678xA SMU must be between 1 and 4 feet.
A.
+S + -S
B.
Current
shunt
+S + -S
+
Agilent 3458A
configured as a
voltmeter
D.
+S + -S
+
Agilent 3458A
configured as
an ammeter
Agilent 3458A
configured as a
voltmeter
+
+
E.
Aux Meas
+
Aux Meas
-
+
DC
voltmeter
F.
+
Electronic load
- - G
+ +
+s G -s
sense jumpers
C.
25 Ohm resistor
(not used for Models
N6753A & N6754A)
+S + -S
+S +LS -LS -S
External
DC supply
Current
monitor
+S + -S
+
G.
+
Agilent 3458A
configured as a
voltmeter
Agilent 3458A
configured as a
voltmeter
- - G
+ +
+s G -s
sense jumpers