Specifications

Series N6700 Service Guide 43
3
Calibration
Calibration Description ..................................................................................... 44
Calibration Procedure Models N673xB, N674xB, N677xA ..................... 46
Calibration Procedure Models N675xA ...................................................... 49
Calibration Procedure Models N676xA ...................................................... 53
Calibration Procedure Models N678xA SMU ............................................ 61
Calibration Procedure Models N6783A-BDA and -MFG ......................... 75
Change the Admin/Calibration Password .................................................... 79
This chapter includes calibration procedures for the Agilent N6700
Modular Power System. Instructions are given for performing the
procedures from either the front panel or a controller over the GPIB.
Refer to the “Equipment Required” section in chapter 2 for a list of
the equipment required for calibration. Also refer to “Measurement
Techniques” for information about connecting the voltmeter and
current shunt.
NOTE
Perform the verification tests before calibrating your power system. If the power
system passes the verification tests, the unit is operating within its calibration
limits and does not need to be re-calibrated.
Calibration Interval
The recommended calibration interval for Agilent N6700 Series
power modules is one year. Agilent N6700A/B, N6701A, and N6702A
MPS mainframes do not require calibration.