Specifications

2 Performance Verification
36 Series N6700 Service Guide
Test Record Form Agilent N6775A
Agilent Model N6775A Report No _______________ Date __________________
Description Model Minimum Specs. Results Maximum Specs.
Constant Voltage Tests
Voltage Programming & Readback
Minimum Voltage Vout
Voltage measured over interface
High Voltage Vout
Voltage measured over interface
+ 10 mV
Vout 60 mV
59.88 V
Vout 120 mV
__________
__________
__________
__________
+ 130 mV
Vout + 60 mV
60.12 V
Vout + 120 mV
CV Load Effect
24 mV
__________
+ 24 mV
CV Source Effect
6 mV
__________
+ 6 mV
CV Ripple and Noise
peak-to-peak
rms
N/A
N/A
__________
__________
+ 35 mV
+ 9 mV
Transient Response
Voltage at 250 µs
500 mV
__________
+ 500 mV
Constant Current Tests
Current Programming & Readback
Minimum Current Iout
Current measured over interface
High Current Iout
Current measured over interface
52.5 mA
Iout 12 mA
4.9325 A
Iout 19.5 mA
__________
__________
__________
__________
+ 67.5 mA
Iout + 12 mA
5.0675 A
Iout + 19.5 mA
CC Load Effect
6 mA
__________
+ 6 mA
CC Source Effect
1 mA
__________
+ 1 mA
Test Description
N6775A Setting
Voltage Programming & Readback, Min 70 mV, 5 A
Voltage Programming & Readback, High 60 V, 5 A
CV Load Effect, Source Effect, Ripple and Noise 60 V, 5 A
Transient Response 60 V, from 2.5 A to 5 A
Current Programming & Readback, Min 7.5 mA, 60 V
Current Programming & Readback, High 5 A, 60 V
CC Load Effect, Source Effect 5 A, 60 V