Test Record Form Agilent N6736B and N6746B
Agilent N6736B and N6746B Report No _______________ Date __________________
Description Model Minimum Specs. Results Maximum Specs.
Constant Voltage Tests
Voltage Programming & Readback
Minimum Voltage Vout
Voltage measured over interface
High Voltage Vout
Voltage measured over interface
Both
Both
Both
Both
0 mV
Vout 100 mV
99.8 V
Vout 200 mV
__________
__________
__________
__________
+ 200 mV
Vout + 100 mV
100.2 V
Vout + 200 mV
CV Load Effect
N6736B
N6746B
20 mV
30 mV
__________
__________
+ 20 mV
+ 30 mV
CV Source Effect
Both
10 mV
__________
+ 10 mV
CV Ripple and Noise
peak-to-peak
rms
Both
Both
N/A
N/A
__________
__________
+ 30 mV
+ 18 mV
Transient Response
Voltage at 200 µs
N6736B
N6746B
500 mV
1000 mV
__________
__________
+ 500 mV
+ 1000 mV
Constant Current Tests
Current Programming & Readback
Minimum Current Iout
Current measured over interface
High Current Iout
Current measured over interface
Both
Both
N6736B
N6746B
N6736B
N6746B
8.5 mA
Iout 2 mA
0.4893 A
0.9885 A
Iout 2.75 mA
Iout 3.5 mA
__________
__________
__________
__________
__________
__________
+ 11.5 mA
Iout + 2 mA
0.5107 A
1.0115 A
Iout + 2.75 mA
Iout + 3.5 mA
CC Load Effect
Both
2 mA
__________
+ 2 mA
CC Source Effect
Both
1 mA
__________
+ 1 mA
Instrument Settings
Test Description N6736B N6746B
Voltage Programming & Readback, Min Voltage 100 mV, 0.5 A 100 mV, 1 A
Voltage Programming & Readback, High Voltage 100 V, 0.5 A 100 V, 1 A
CV Load Effect, Source Effect, Ripple and Noise
100 V, 0.5 A
100 V, 1 A
Transient Response 100 V, from 0.25 A to 0.5 A
100 V, from 0.5 A to 1 A
Current Programming & Readback, Min Current 1.5 mA, 100 V 1.5 mA, 100 V
Current Programming & Readback, High Current 0.5 A, 100 V 1 A, 100 V