Specifications
Performance Verification 2
Series N6700 Service Guide 19
“Current Readback, 200 μA Current”. Record the current reading
of the Agilent 3458A and the current measured over the
interface. Readings should be within limits specified for the
appropriate model under “Current Readback, 200μA Range”.
8 For models N6763A–N6766A, and N6783A only. Set the current
measurement to the low range. Set the instrument as described
under “Current Readback, Low Current”. Repeat step 5. Record
the current measured over the interface. The readings should be
within the limits specified under “Low Current Readback Low
Range”.
9 Steps 9 through 13 for models N678xA SMU only.. Connect the
Agilent 3458A directly across the output (see Test Setup D
without the resistor). Set the Agilent 3458A to the 100 mA range
(no autoranging). Set the current programming to the low ranges
- 300 mA for N6781A and N6782A; 100 mA for the N6784A. Set
the instrument as described under “Current Programming 300
mA & 100 mA ranges”. Record the current readings of the 3458A.
The readings should be within the limits specified under “300 mA
and 100 mA range Current Iout”.
10 Set the current programming to the 10 mA range for the N6784A.
Set the instrument as described under “Current Programming 10
mA range”. Record the current reading of the 3458A. The reading
should be within the limits specified under “10 mA range Current
Iout”.
11 Connect the Agilent 3458A in series with a 10 Ω resistor to the
output (see Test Setup D). Set the Agilent 3458A to measure
current using the 100 mA range. Set the current measurement to
the 100 mA range. Set the instrument as described under
“Current Readback 100 mA Range”. Record the current reading
of the 3458A and the current measured over the interface. The
readings should be within the limits specified under “100 mA
Range Readback”.
12 Connect the Agilent 3458A in series with a 10 kΩ resistor to the
output (see Test Setup D). Set the Agilent 3458A to measure
current using the 10 mA range. Set the current measurement to
the 1 mA range. Set the instrument as described under “Current
Readback 1 mA Range”. Record the current reading of the 3458A
and the current measured over the interface. The readings should
be within the limits specified under “1 mA Range Readback”.
13 Connect the Agilent 3458A in series with a 1 MΩ resistor to the
output (see Test Setup D). Set the Agilent 3458A to measure
current using the 100 μA range. The remote command is “DCI
100e-6”. Set the current measurement to the 10 μA range. Set the
instrument as described under “Current Readback, 10 μA Range”.
Record the current reading of the 3458A and the current
measured over the interface. The readings should be within the
limits specified under “10 μA Range Readback”.