Menu Command Control Description
System IO DigPort Pin1 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, FaultOut.
Specifies the pin polarity: Positive, Negative
Pin2 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut.
Specifies the pin polarity: Positive, Negative
Pin3 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, InhibitIn.
Specifies the pin polarity: Positive, Negative
Pin4 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.
Specifies the pin polarity: Positive, Negative
Pin5 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.
Specifies the pin polarity: Positive, Negative
Pin6 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.
Specifies the pin polarity: Positive, Negative
Pin7 Function
Polarity
Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.
Specifies the pin polarity: Positive, Negative
Data Sends/reads data from the digital I/O port
Groups Defines groups of output channels that are connected in parallel.
Preferences Display Contrast Configures the display contrast.
Saver Configures the screen saver and wake-on I/O timer.
View Selects 1-channel or 4-channel view at turn-on
Keys Enables/disables key clicks and configures the On/Off key.
Lock Locks front panel keys. Enter a password to unlock the front panel.
Admin Login/Logout Enter a password to access the admin functions.
Cal Function Vprog High Enters measured data for the High calibration point.
Low Enters measured data for the Low calibration point.
Vmeas Enters measured data.
CMRR Calibrates common mode rejection ratio.
Iprog High Enters measured data for the High calibration point.
Low Enters measured data for the Low calibration point.
Imeas 100mA Enters measured data for the 100 mA range point.
200uA Enters measured data for the 200 ยตA range point.
Dprog
Calibrates the downprogrammer.
Ipeak Calibrates I peak.
Date Saves the calibration date for each channel.
Save Saves the calibration data.
LAN Enables/disables the LAN interface and the built-in Web server.
USB Enables/disables the USB interface.
Nvram Resets all non-volatile RAM settings to their factory defaults.
Password
Changes the password for the admin functions.
About Frame Displays model, serial number, and firmware revisions.
Module Displays model, serial number, options, voltage, current, power.