Programming instructions
Table Of Contents
- Programming Guide
- Table of Contents
- 1 Getting Started
- 2 Programming Examples
- Using the Programming Examples
- GPIB Programming Examples
- Before Using the Examples
- Interface Check using Agilent BASIC
- Interface Check Using NI-488.2 and C++
- Interface Check using VISA and C
- Local Lockout Using Agilent BASIC
- Local Lockout Using NI-488.2 and C++
- Queries Using Agilent BASIC
- Queries Using NI-488.2 and C++
- Queries Using VISA and C
- Generating a CW Signal Using VISA and C
- Generating an Externally Applied AC-Coupled FM Signal Using VISA and C
- Generating an Internal AC-Coupled FM Signal Using VISA and C
- Generating a Step-Swept Signal Using VISA and C
- Saving and Recalling States Using VISA and C
- Reading the Data Questionable Status Register Using VISA and C
- Reading the Service Request Interrupt (SRQ) Using VISA and C
- LAN Programming Examples
- RS-232 Programming Examples
- 3 Programming the StatusRegisterSystem
- 4 Command Reference
- Command Reference Information
- SCPI Basics
- IEEE 488.2 Common Commands
- Calibration subsystem
- Communication Subsystem
- Diagnostic Subsystem
- Display Subsystem
- Memory Subsystem
- Mass Memory Subsystem
- Output Subsystem
- Status Subsystem
- :OPERation:CONDition
- :OPERation:ENABle
- :OPERation:NTRansition
- :OPERation:PTRansition
- :OPERation[:EVENt]
- :PRESet
- :QUEStionable:CALibration:CONDition
- :QUEStionable:CALibration:ENABle
- :QUEStionable:CALibration:NTRansition
- :QUEStionable:CALibration:PTRansition
- :QUEStionable:CALibration[:EVENt]
- :QUEStionable:CONDition
- :QUEStionable:ENABle
- :QUEStionable:FREQuency:CONDition
- :QUEStionable:FREQuency:ENABle
- :QUEStionable:FREQuency:NTRansition
- :QUEStionable:FREQuency:PTRansition
- :QUEStionable:FREQuency[:EVENt]
- :QUEStionable:MODulation:CONDition
- :QUEStionable:MODulation:ENABle
- :QUEStionable:MODulation:NTRansition
- :QUEStionable:MODulation:PTRansition
- :QUEStionable:MODulation[:EVENt]
- :QUEStionable:NTRansition
- :QUEStionable:POWer:CONDition
- :QUEStionable:POWer:ENABle
- :QUEStionable:POWer:NTRansition
- :QUEStionable:POWer:PTRansition
- :QUEStionable:POWer[:EVENt]
- :QUEStionable:PTRansition
- :QUEStionable[:EVENt]
- System Subsystem
- Trigger Subsystem
- Unit Subsystem (:UNIT)
- Amplitude Modulation Subsystem
- :AM[1]|2...
- :AM:INTernal:FREQuency:STEP[:INCRement]
- :AM:MODE
- :AM[1]|2:EXTernal[1]|2:COUPling
- :AM[1]|2:EXTernal[1]|2:IMPedance
- :AM[1]|2:INTernal[1]:FREQuency:ALTernate
- :AM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent
- :AM[1]|2:INTernal[1]:SWEep:RATE
- :AM[1]|2:INTernal[1]:SWEep:TRIGger
- :AM[1]|2:INTernal[1]|2:FREQuency
- :AM[1]|2:INTernal[1]|2:FUNCtion:NOISe
- :AM[1]|2:INTernal[1]|2:FUNCtion:RAMP
- :AM[1]|2:INTernal[1]|2:FUNCtion:SHAPe
- :AM[1]|2:SOURce
- :AM[1]|2:STATe
- :AM[1]|2:TYPE
- :AM[1]|2[:DEPTh]:EXPonential
- :AM[1]|2[:DEPTh][:LINear]
- :AM[1]|2[:DEPTh][:LINear]:TRACk
- :AM[:DEPTh]:STEP[:INCRement]
- Correction Subsystem ([:SOURce]:CORRection)
- Frequency Subsystem ([:SOURce])
- Frequency Modulation Subsystem ([:SOURce])
- :FM[1]|2...
- :FM:INTernal:FREQuency:STEP
- :FM[1]|2:EXTernal[1]|2:COUPLing
- :FM[1]|2:EXTernal[1]|2:IMPedance
- :FM[1]|2:INTernal[1]:FREQuency:ALTernate
- :FM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent
- :FM[1]|2:INTernal[1]:SWEep:RATE
- :FM[1]|2:INTernal[1]:SWEep:TRIGger
- :FM[1]|2:INTernal[1]|2:FREQuency
- :FM[1]|2:INTernal[1]|2:FUNCtion:NOISe
- :FM[1]|2:INTernal[1]|2:FUNCtion:RAMP
- :FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe
- :FM[1]|2:SOURce
- :FM[1]|2:STATe
- :FM[1]|2[:DEViation]
- :FM[1]|2[:DEViation]:TRACk
- List/Sweep subsystem ([:SOURce])
- Low Frequency Output Subsystem ([:SOURce]:LFOutput)
- :AMPLitude
- :FUNCtion[1]:FREQuency:ALTernate
- :FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent
- :FUNCtion[1]:SWEep:RATE
- :FUNCtion[1]:SWEep:TRIGger
- :FUNCtion[1]|2:FREQuency
- :FUNCtion[1]|2:SHAPe
- :FUNCtion:NOISe
- :FUNCtion[1]|2:SHAPe:RAMP
- :SOURce
- LF Out softkeys:LF Out Off On;low frequency output subsystem keys:LF Out Off On
- Phase Modulation subsystem
- :PM[1]|2...
- :PM:INTernal:FREQuency:STEP[:INCRement]
- :PM[1]|2:BANDwidth|BWIDth
- :PM[1]|2:EXTernal[1]|2:COUPling
- :PM[1]|2:EXTernal[1]|2:IMPedance
- :PM[1]|2:INTernal[1]:FREQuency:ALTernate
- :PM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent
- :PM[1]|2:INTernal[1]:SWEep:RATE
- :PM[1]|2:INTernal[1]:SWEep:TRIGger
- :PM[1]|2:INTernal[1]|2:FREQuency
- :PM[1]|2:INTernal[1]|2:FUNCtion:NOISe
- :PM[1]|2:INTernal[1]|2:FUNCtion:RAMP
- :PM[1]|2:INTernal[1]|2:FUNCtion:SHAPe
- :PM[1]|2:SOURce
- :PM[1]|2:STATe
- :PM[1]|2[:DEViation]
- :PM[1]|2[:DEViation]:TRACk
- :PM[:DEViation]:STEP[:INCRement]
- Power Subsystem ([:SOURce])
- :POWer:ALC:BANDwidth|BWIDth
- :POWer:ALC:BANDwidth|BWIDth:AUTO
- :POWer:ALC:LEVel
- :POWer:ALC:SEARch
- :POWer:ALC:SOURce
- :POWer:ALC:SOURce:EXTernal:COUPling
- :POWer:ALC[:STATe]
- :POWer:ATTenuation
- :POWer:ATTenuation:AUTO
- :POWer:MODE
- :POWer:REFerence
- :POWer:REFerence:STATe
- :POWer:STARt
- :POWer:STOP
- :POWer[:LEVel][:IMMediate]:OFFSet
- :POWer[:LEVel][:IMMediate][:AMPLitude]
- Pulse Modulation Subsystem ([:SOURce])
- SCPI Command Compatibility
- Index

Command Reference
Pulse Modulation Subsystem ([:SOURce])
Chapter 4290
:PULM:INTernal[1]:PERiod
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:PERiod <val><unit>|UP|DOWN
[:SOURce]:PULM:INTernal[1]:PERiod?
This command sets the period for the internally generated pulse modulation source.
*RST
+
2.00000000E
−
006
Range 70nS–42S
Key Entry
Pulse Period
Remarks If the entered value for the pulse period is equal to or less than the
value for the pulse width, the pulse width changes to a value that is
less than the pulse period.
Refer to “:PULM:INTernal[1]:PERiod:STEP[:INCRement]” for setting
the value associated with the UP and DOWN choices.
:PULM:INTernal[1]:PERiod:STEP[:INCRement]
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:PERiod:STEP[:INCRement] <val><unit>
[:SOURce]:PULM:INTernal[1]:PERiod:STEP[:INCRement]?
This command sets the step increment for the internal pulse period.
*RST +1.00000000E-006
Range 10nS–42S
Key Entry N/A
Remarks The value set by this command is used with the UP and DOWN choices
for the pulse period command. Refer to “:PULM:INTernal[1]:PERiod”
for more information.
Choices <val><unit> UP DOWN