Programming instructions

Table Of Contents
Command Reference
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
Chapter 4264
:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent
Supported PSG-A Series
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:
PERCent <val><unit>
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent?
This command sets the amplitude of the second tone for a dual-sine waveform as a
percentage of the total LF output amplitude. For example, if the second tone makes up
30% of the total amplitude, then the first tone is 70% of the total amplitude.
*RST
+
5.00000000E
+
001
Range 0–100PCT
Key Entry
LF Out Tone 2 Ampl % of Peak
Remarks Refer to “:FUNCtion[1]|2:SHAPe” on page 266 for selecting the
waveform type.
:FUNCtion[1]:SWEep:RATE
Supported PSG-A Series
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:RATE <val><unit>
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:RATE?
This command sets the sweep rate for an internally generated swept-sine signal at the
LF output.
The variable <val> has a minimum resolution of 0.5 hertz.
*RST
+
4.00000000E
+
002
Range 0.5HZ–100kHZ
Key Entry
LF Out Sweep Rate
Remarks Refer to “:FUNCtion[1]|2:SHAPe” on page 266 for selecting the
waveform type.