Status Subsystem
 This subsystem programs the power supply status registers. The power supply has three groups of status registers;
Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as
described in "Chapter 4 - Status Reporting". The Operation and Questionable status groups each consist of the Condition,
Enable, and Event registers and the NTR and PTR filters.
STAT:PRES
 This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsystem NTR and
Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.
  Command Syntax STATus:PRESet
  Parameters (None)
  Examples STAT:PRES STATUS:PRESET
  Query Syntax (None)
  Related Commands (None)
Status Operation Registers
 The bit configuration of all Status Operation registers is shown in the following table:
 Bit Configuration of Operation Registers
 Bit Position
 15-12  11  10  9  8  7  6  5  4  3  2  1  0
 Bit Name
 NU  NU  CC  NU  CV  NU  NU  WTG  NU  NU  NU  NU  CAL
 Bit Weight
 2048  1024  512  256  128  64  32  16  8  4  2  1
 CAL = Interface is computing new calibration constants; CC = The power module is in constant current mode.
 CV = The power module is in constant voltage mode; NU = (Not used); WTG = Interface is waiting for a trigger.
Note See "Chapter 4 - Status Reporting" for more explanation of these registers.
STAT:OPER?
 This query returns the value of the Operation Event register. The Event register is a read-only register which holds (latches)
all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.
  Query Syntax  STATus:OPERtion[:EVENt]?
  Parameters (None)
  Returned Parameters <NR1> (Register Value)
  Examples STAT:OPER? STATUS:OPERATIOBAL:EVENT?
  Related Commands  *CLS STAT:OPER:NTR STAT:OPER:PTR
STAT:OPER:COND?
 This query returns the value of the Operation Condition register. That is a read-only register which holds the real-time
(unlatched) operational status of the power supply.
  Query Syntax STATus:OPERation:CONDition?
  Parameters (None)
  Examples STAT:OPER:COND? STATUS:OPERATION:CONDITION?
  Returned Parameters <NR1> (Register value)
  Related Commands (None)