User manual
 Programming the DC Source - 3
25
Internally Triggered Measurements
 You can use the data acquisition trigger system to synchronize the timing of the voltage and current data
acquisition with a BUS or internal trigger source. Then use the FETCh commands to return different
calculations from the data acquired by the measurement trigger.
 SCPI Triggering Nomenclature
 As previously explained under "Triggering Output Changes", the dc source uses the following sequence
name and alias for the measurement trigger system. This alias can be used instead of the sequence form.
 Sequence Form  Alias
 SEQuence2  ACQuire
 Measurement Trigger System Model
 Figure 3-2 is a model of the measurement trigger system. The rectangular boxes represent states. The
arrows show the transitions between states. These are labeled with the input or event that causes the
transition to occur.
ACQUIRED
INITIATED STATE
IDLE STATE
ABORt
*RCL
*RST
TRIGGER RECEIVED
INITiate
[
:IMMediate
]
TRIG
g
er:COUNt
COMPLETE?
YES
NO
SENSe:SWEep:POINts
 Figure 3-2. Model of Measurement Triggers
 Initiating the Measurement Trigger System (Agilent 66312A, 66332A Only)
 When the dc source is turned on, the trigger system is in the idle state. In this state, the trigger system
ignores all triggers. Sending the following commands at any time returns the trigger system to the Idle
state:
ABORt
*RST
*RCL
 The INITiate commands move the trigger system from the Idle state to the Initiated state. This enables
the dc source to receive triggers. To initiate for a measurement trigger, use: