Specifications
13
Page 25
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Comparing the 8510 and PNA
8510 (85108A)
• Dominant mode is wideband detection
• Detection is done BEFORE analog-to-digital conversion
• Analog synchronous detector produces baseband I/Q output
(detector bandwidth = 1.5 MHz)
• Pulse profiling achieved by varying sample point of baseband pulses
• Trade off speed and dynamic range with averaging
PNA
• Dominant mode is narrowband detection
• All processing (filtering and detection) is done digitally
• Widest bandwidth = 35 kHz
• Pulse profiling achieved with analog switches that gate IF (or RF) signals
• Trade off speed and dynamic range with variable IF bandwidths and averaging
PNA-L
• Dominant mode is wideband detection
• All processing (filtering and detection) is done digitally
• Widest bandwidth = 600 kHz
• Trade off speed and dynamic range with variable IF bandwidths and averaging
Page 26
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Agilent TR Module Test Systems
85120A S10 Family