Receiver calibration
Receiver calibration gives corrected absolute power reading (for
unratioed traces as opposed to an S-parameter traces)
Often used with FOM for measurements like harmonics, TOI
Uses power-meter-corrected internal source as a reference
(source power cal)
Now saved as a cal set
Multiple receiver cals requires
separate channels for each cal
Source power calibrations
No change from current behavior
Saved as part of instrument states
Not saved in cal sets
New!
Page 18
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Correction Indicator on Status Bar
C - Full correction
“C” is displayed immediately after a calibration is performed
or when a valid Cal Set is applied. If you require optimum
accuracy, avoid adjusting the analyzer’s settings after calibration.
C* - Interpolated correction
"C*" appears in the status bar when interpolation is used during a measurement.
CChanged settings
"C" appears in the status bar when one or more of the following stimulus settings change. The
resulting measurement accuracy depends on which parameter has changed and how much it has
changed. For optimum accuracy, recalibrate using the new settings.
•Sweep time
• IF bandwidth
•Port power
• Stepped sweep enabled/disabled
LowestNo correctionNo Cor
UncertainChangedC
UncertainInterpolatedC*
HighestFullC
AccuracyCorrection levelSymbol